A Polypyrrole-based Strain Sensor Dedicated to Measure Bladder Volume in Patients with Urinary Dysfunction

Sumitra Rajagopalan, Mohamad Sawan, Ebrahim Ghafar‐Zadeh, O. Savadogo, Vamsy P. ChodavarapuSensors 2008 · DOI

TL;DR

This paper introduces a novel polypyrrole-coated elastic fabric strain sensor designed to measure bladder volume in patients with urinary dysfunction. By leveraging the material's conductivity and elasticity, the sensor offers a non-invasive and potentially implantable solution for monitoring bladder fullness, addressing a critical need for restoring sensory feedback and voluntary control.

What Problem Does This Paper Address?

Urinary bladder dysfunction affects millions, leading to loss of voluntary control and crucial sensory feedback about bladder fullness. Current methods like catheterization are invasive and carry infection risks. While neural stimulation can restore control, it doesn't address the sensory feedback gap. Existing non-invasive or implantable devices for bladder volume assessment have limitations, such as being invasive, having limited accuracy, or not effectively measuring intermediate volumes. This paper aims to fill this gap by developing a sensor that can accurately and reliably measure bladder volume, enabling patients to regain awareness of bladder fullness and trigger voiding when necessary.

Methods

The core innovation is a strain sensor fabricated by coating a highly elastic fabric with polypyrrole (PPY), an electronically conducting polymer. This PPY-coated fabric is designed to be placed around the bladder, acting like a snug pouch. As the bladder fills and stretches, the fabric deforms, causing a change in the electrical resistance of the PPY layer. This change in resistance is then correlated to the bladder volume.

Key methodological aspects include:

  • Material Selection: A highly elastic fabric (nylon hosiery) was chosen for its ability to mimic the viscoelastic properties of bladder tissue.
  • Polypyrrole Deposition: Two methods were explored: solution-phase and vapor-phase polymerization. Vapor-phase polymerization was favored for producing a more uniform, stable, and reproducible coating.
  • Sensor Design: The sensor is envisioned as a pouch that fits over the bladder, minimizing the need for extensive surgical sutures.
  • Measurement Principle: The sensor relies on the piezoresistive effect of polypyrrole, where mechanical strain (stretching) leads to a change in electrical resistance.
  • Circuitry: An implantable interface circuit (shown in Figure 1c) is proposed for reading out the resistance changes, utilizing a resistor-less technique for improved accuracy.
  • Mechanical Characterization: Uniaxial tensile properties of the PPY-coated fabric were tested using a mechanical analyzer to compare its elasticity with bladder tissue.
  • In-Vitro Testing: The sensor was tested on a rubber balloon simulating a bladder, with water gradually added to measure resistance changes at different volumes.

Datasets

Not discussed in the paper. The paper focuses on the development and characterization of a novel sensor and its in-vitro testing on a phantom bladder.

Key Insights and Ideas

  • Elasticity is Key: The paper highlights that the bladder wall is highly viscoelastic and can stretch significantly. Therefore, a sensor that is also highly elastic and responsive to strain is crucial for accurate volume measurement, unlike rigid sensors.
  • PPY as a Smart Material: Polypyrrole's unique combination of electrical conductivity and flexibility makes it an ideal candidate for this application. Its resistance changes predictably with mechanical deformation.
  • Fabrication Matters: The choice of fabrication method (vapor-phase polymerization) significantly impacts the sensor's uniformity, stability, and reproducibility, which are critical for reliable biomedical devices.
  • Pouch Design Advantage: The proposed pouch design offers a significant advantage for implantable applications, as it can conform to the bladder's shape and potentially be implanted with minimal surgical intervention, accommodating the irregular shapes of dysfunctional bladders.
  • Passive Interrogation Potential: The authors suggest that with chemical modifications to decrease resistance, the sensor could be interrogated passively using resonance circuits, eliminating the need for implanted batteries and circuits, which is a major step towards truly unobtrusive implants.

Key Results

  • The PPY-coated fabric exhibits a non-linear stress-strain behavior comparable to bladder tissue, with a Young's modulus of 25, making it suitable for mimicking bladder elasticity.
  • The sensor demonstrated a reproducible change in electrical resistance upon stretching.
  • The resistance response to stretching was found to be linear in the 20%-40% strain variation range.
  • In-vitro testing on a phantom bladder showed an approximately linear resistance response in the 40-100 ml range and a linear response and sensitivity in the 100-300 ml range, which is relevant for normal micturition reflexes.
  • Vapor-phase polymerization yielded a more uniformly coated polypyrrole fabric that remained stable over many stress-strain cycles.

Limitations and Future Work

Limitations:

  • The study primarily focused on in-vitro testing using a phantom bladder.
  • The paper acknowledges that the resistance change is influenced by chemical fabrication conditions, implying that further optimization is needed.
  • While the pouch design is proposed to minimize sutures, the actual surgical implantation procedure and its ease are not fully demonstrated.

Future Work:

  • In-vivo testing: The most critical next step is to conduct in-vivo testing to validate the sensor's performance in a biological environment.
  • Fabrication Optimization: Further research is needed to optimize fabrication techniques and chemical modifications to enhance sensor stability, sensitivity, conductivity, and reversibility.
  • Integrated System Development: The authors propose fabricating an interface chip (as shown in Figure 1c) and developing a fully functioning system, including wireless transmission to a wearable display.
  • Passive Interrogation: Exploring chemical modifications to enable passive interrogation of the sensor would be a significant advancement, potentially eliminating the need for implanted electronics.
  • Addressing Dysfunctional Bladder Shapes: The sensor's ability to accommodate the irregular shapes of dysfunctional bladders needs further in-vivo validation.

Summary updated

Abstract

This paper describes a new technique to measure urine volume in patients with urinary bladder dysfunction. Polypyrrole - an electronically conducting polymer - is chemically deposited on a highly elastic fabric. This fabric, when placed around a phantom bladder, produced a reproducible change in electrical resistance on stretching. The resistance response to stretching is linear in 20%-40% strain variation. This change in resistance is influenced by chemical fabrication conditions. We also demonstrate the dynamic mechanical testing of the patterned polypyrrole on fabric in order to show the feasibility of passive interrogation of the strain sensor for biomedical sensing applications.