A one-dimensional model for aspiration therapy in blood vessels

Michaël Herty, Niklas Kolbe, Michael NeidlinarXiv 2024 · DOI · arXiv

Abstract

Aspiration thrombectomy is a treatment option for ischemic stroke due to occlusions in large vessels. During the therapy a device is inserted into the vessel and suction is applied. A new one-dimensional model is introduced that is capable of simulating this procedure while accounting for the fluid-structure interactions in blood flow. To solve the coupling problem at the tip of the device a problem-suited Riemann solver is constructed based on relaxation of the hyperbolic model. Numerical experiments investigating the role of the catheter size and the suction forces are presented.

Introduction

Ischemic strokes, often caused by blood clots (occlusions) in large brain vessels, are a leading cause of disability worldwide. A common treatment is aspiration thrombectomy, where a distal catheter is navigated to the clot to remove it via suction. To plan these interventions effectively, clinicians need to understand the complex fluid dynamics involved.

While 3D Computational Fluid Dynamics (CFD) provides high detail, it is computationally expensive and often ignores the elasticity of the blood vessel walls (Fluid-Structure Interaction, or FSI). Conversely, standard 1D models are fast but have historically failed to account for the physical presence of the catheter or the specific flow physics occurring at the catheter's tip. This research introduces a 1D hyperbolic model that bridges this gap, offering a computationally efficient tool for simulating aspiration therapy.

Methods

The proposed model is based on a 1D hyperbolic system of partial differential equations (PDEs) for mass conservation and momentum balance. It incorporates several key technical components:

  • Fluid-Structure Interaction (FSI): The model uses an algebraic relationship between vessel pressure and cross-sectional area, accounting for the Young modulus and wall thickness of the vessel.
  • Catheter Representation: The catheter is treated as a rigid pipe inserted into the elastic vessel. This creates a "catheterized" segment where the available area for blood flow is the vessel area minus the catheter area.
Schematic of the 1D catheter model geometry, showing the vessel area (A), catheter area (Ac), and flow velocities (u, w).
  • The Coupling Problem at the Tip: The transition from the free vessel to the catheterized segment is a point of discontinuity. To solve the equations at this interface, a problem-specific Riemann solver was constructed using a "Jin/Xin relaxation" approach. This ensures continuity of the gross vessel area and average flow velocity across the tip, even during active suction.

Results and Key Insights

The model was validated using in-silico numerical experiments with physiological parameters. The results highlight several critical dynamics of the procedure:

Insertion and Suction Effects

Simply inserting the catheter creates a forward-propagating shockwave that decreases pressure and reduces flow rate, an effect that magnifies with larger catheter diameters. When strong suction is applied, the model captures a "kink-shaped" minimum in the flow rate at the tip position.

Numerical simulation results showing flow rate (Q), pressure (p), and vessel area (A) at different time steps (t=0, 0.005, 0.01).

Vessel Constriction

A major insight is the model's ability to capture vessel constriction. High suction forces can cause the elastic vessel wall to narrow significantly or even collapse at the tip. This is a critical factor for surgical safety that 1D models previously struggled to represent.

Occlusion Mitigation

In simulated stroke scenarios, blood clots cause high-pressure waves to reflect back into the system. The model demonstrates that activating aspiration successfully reduces this high-pressure period, potentially protecting the vessel from damage.

Discussion and Future Work

By reducing the problem to 1D, this model can simulate complex scenarios—such as a full heartbeat with an occlusion—much faster than 3D CFD, making it a potential candidate for real-time clinical decision support.

However, limitations remain, including the assumption of a flat velocity profile and a perfectly rigid catheter. Future work will focus on integrating this catheter model into full 1D networks, such as the Circle of Willis, to better represent the complex branching structure of the human brain's vasculature.

Summary updated

Abstract

Aspiration thrombectomy is a treatment option for ischemic stroke due to occlusions in large vessels. During the therapy a device is inserted into the vessel and suction is applied. A new one-dimensional model is introduced that is capable of simulating this procedure while accounting for the fluid-structure interactions in blood flow. To solve the coupling problem at the tip of the device a problem-suited Riemann solver is constructed based on relaxation of the hyperbolic model. Numerical experiments investigating the role of the catheter size and the suction forces are presented.
BibTeX
@misc{herty2024one,
  author = {Michaël Herty and Niklas Kolbe and Michael Neidlin},
  title = {{A one-dimensional model for aspiration therapy in blood vessels}},
  year = {2024},
  eprint = {2403.05494},
  archivePrefix = {arXiv},
  url = {https://arxiv.org/abs/2403.05494},
}

Loading related content...

Figure