A novel method for assessing position-sensitive detector performance

N.h. Clinthorne, W.l. Rogers, Ling Shao, Alfred O. Hero, Kenneth F. KoralIEEE Transactions on Nuclear Science 1989 · DOI

Abstract

A marked-point process model of a position-sensitive detector is developed which includes the effects of detector efficiency, spatial response, energy response, and source statistics. The average mutual information between the incident distribution of gamma -rays and the detector response is derived and used as a performance index of detector optimization. A brief example is presented which uses this figure-of-merit for optimization of light-guide dimensions for the SPRINT IT modular scintillation camera.< <ETX xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">&gt;</ETX>
BibTeX
@article{clinthorne1989novel,
  author = {N.H. Clinthorne and W.L. Rogers and Ling Shao and Alfred O. Hero and Kenneth F. Koral},
  title = {{A novel method for assessing position-sensitive detector performance}},
  journal = {IEEE Transactions on Nuclear Science},
  year = {1989},
  doi = {10.1109/23.34617},
}