A novel instrument guidance device for enhanced surgical navigation

Xuelong Wang, Changjiang Zhou, Zheng‐Yu Huang, Hao Zhao, Shiju YanProgress 2025 · DOI

Abstract

The key to an ultrasound-guided system is to accurately determine the spatial position of the surgical instrument. Traditional optical positioning devices often restrict a surgeon's movements and occupy significant operating room space. This paper presents a compact device designed to provide enhanced navigation information. Equipped with an angle sensor and a manual depth-adjustment button, the device measures the relative angle between a puncture needle and an ultrasound probe in real time. A custom two-stage filtering algorithm was developed to reduce signal noise. Results show significant improvements in signal smoothness and a low average point-to-point error of 3.53 pixels, demonstrating the device's potential to support precise surgical operations without the need for bulky external hardware.

Introduction

Ultrasound-guided surgery is a cornerstone of minimally invasive procedures, yet maintaining spatial awareness of the needle's path relative to the probe remains a significant challenge. Current "gold standard" solutions typically rely on optical tracking systems. These systems require mounting bulky targets on surgical tools and positioning binocular cameras around the operating room, which can obstruct the surgeon’s line-of-sight and consume valuable space.

To address these limitations, researchers have developed a mechatronic attachment that "shrinks" the navigation system. By integrating sensors directly onto the ultrasound probe, the device provides real-time guidance and depth estimation without the need for external tracking hardware.

System Design and Hardware Implementation

The proposed device is a 3D-printed mechatronic sleeve designed to fit standard ultrasound probes. Its primary function is to measure the relative tilt between the puncture needle and the probe using a high-precision RDC501051A resistive angle sensor.

Hardware components (A) including the PCB and buttons, and the fully assembled device attached to an ultrasound probe (B).

The system's core is an STM32F103C8T6 microcontroller, which processes data from the angle sensor via a 16-bit ADS1110 analog-to-digital converter (ADC). This architecture ensures that even minute changes in the needle's orientation are captured with high fidelity, allowing for precise tracking of the instrument's trajectory.

The electronic schematic of the guidance system, highlighting the STM32 microcontroller and power management circuits.

Signal Processing and Depth Estimation

A critical challenge in medical mechatronics is signal noise caused by electromagnetic interference and mechanical vibrations. To ensure a stable visual guide for the surgeon, a two-stage hybrid software filter was implemented:

  1. Windowed Mean Filter: This stage removes extreme values from a 10-point window and calculates the mean to eliminate transient spikes.
  2. Standard Deviation Outlier Removal: A larger 100-point window is used to identify and strip outliers based on standard deviation, followed by a final mean filter for maximum smoothness.

Furthermore, the device includes a physical button that allows surgeons to adjust a "virtual guide line" on the ultrasound display. This enables pre-puncture depth estimation, established through a polynomial fitting model ($y = 16.32x - 1.074$), allowing for a visual "dry run" of the needle path before the actual procedure begins.

Performance Evaluation and Results

The device was evaluated using a breast body model and a dedicated experimental bench. The filtering algorithm proved highly effective, reducing the Signal Smoothness metric from 0.036989 to 0.0010376 and cutting the Count of Local Extrema (sharp, erratic signal jumps) from 251 to 135.

In accuracy tests, the average point-to-point error between the predicted guide line and the actual needle tip was found to be 3.53 pixels. Depth measurement precision was equally impressive, with an average error of just 0.216 mm and a near-perfect R-squared value of 1 for the polynomial model, indicating a highly reliable correlation between the device's sensor data and physical depth.

Conclusion

The developed guidance device and filtering algorithm provide a compact, accurate, and non-obtrusive alternative to traditional optical tracking systems. By moving the navigation intelligence directly onto the probe, the system supports surgical precision while streamlining the operating room environment. Future work will focus on automating depth measurement using magnetic sensors and increasing the sampling rate to further enhance real-time responsiveness.

Summary updated

Abstract

Objectives: The key to an ultrasound-guided system is to accurately determine the spatial position of the surgical instrument. The size and shape of the optical positioning target in the optical positioning device can restrict the doctor's hand movements. Additionally, the use of a binocular camera occupies valuable operating room space. To reduce both the space occupation and the impact on surgical procedures, this paper presents a device designed to provide enhanced navigation information to assist the surgeon. Methods: The proposed device is equipped with an angle sensor and a button. The angle sensor measures the relative angle between the puncture device and the ultrasound probe in real time. The button assists the surgeon in measuring the depth of the surgical target before performing the puncture. Furthermore, we propose a filtering algorithm to reduce noise in the angle sensor's output signal. The device's performance was evaluated using an ultrasound instrument and an experimental setup for data collection and testing. Results: The filtering algorithm effectively reduced noise in the angular data. After filtering, the Signal Smoothness decreased from 0.036989 to 0.0010376, the Coefficient of Variation decreased from 0.00046682 to 0.0004323, and the Count of Local Extrema decreased from 251 to 135. Additionally, we collected and compared the coordinates of the guide line endpoints at the tip of the puncture needle and found an average point-to-point error of 3.53 pixel. Conclusions: The developed guidance device and filtering algorithm provide valuable navigation data, supporting the surgeon's operations.
BibTeX
@inproceedings{wang2025novel,
  author = {Xuelong Wang and Changjiang Zhou and Zheng‐Yu Huang and Hao Zhao and Shiju Yan},
  title = {{A novel instrument guidance device for enhanced surgical navigation}},
  booktitle = {Progress},
  year = {2025},
  doi = {10.61189/389587wfmupp},
}
Figure