A Novel Design of Water-Activated Variable Stiffness Endoscopic Manipulator with Safe Thermal Insulation

Qian Gao, Zhenglong SunActuators 2021

TL;DR

This paper introduces a novel endoscopic manipulator for Natural Orifice Transluminal Endoscopic Surgery (NOTES) that can dynamically change its stiffness. By using a temperature-sensitive bioplastic activated by hot and cold water, the manipulator can switch between a flexible state for navigation and a rigid state for surgical tasks, all while incorporating a safe thermal insulation system.

What Problem Does This Paper Address?

Natural Orifice Transluminal Endoscopic Surgery (NOTES) offers significant advantages like reduced patient trauma and faster recovery. However, a key challenge in NOTES is the need for an endoscopic tool that can navigate through the body's tortuous natural passages (requiring flexibility) and then become sufficiently rigid to act as a stable platform for surgical instruments. Traditional endoscopes struggle to achieve this dual requirement, often being either too flexible or too rigid. Existing variable stiffness solutions face issues like bulky designs, insufficient rigidity, or safety concerns related to their activation methods (e.g., electric heating). This paper aims to address these limitations by proposing a safer, more effective, and compact variable stiffness endoscopic manipulator.

Methods

The core innovation lies in the material and activation mechanism. The researchers utilize a starch-based bioplastic called FORMcard, which exhibits tunable stiffness based on temperature. Specifically, when heated above its glass transition temperature (around 65°C), it becomes soft and flexible; when cooled below this temperature, it hardens.

The activation process is water-based:

  • Flexible Mode: Hot water (65°C) is pumped through internal channels, heating the FORMcard and rendering the manipulator flexible.
  • Rigid Mode: Cold water (5°C) is pumped through the same channels, cooling the FORMcard and restoring rigidity.

To ensure patient safety, a microencapsulated phase change material (MEPCM) layer is integrated as thermal insulation. This MEPCM, containing paraffin with a melting point around body temperature (37°C), absorbs excess heat from the hot water, preventing the manipulator's surface from reaching dangerously high temperatures and causing burns. The manipulator's structure incorporates two hollow Variable Stiffness Channels (VSCs) made of FORMcard, running through disks that allow for bending and actuation via cables.

Datasets

Not discussed in the paper. The study focuses on experimental validation of a prototype.

Key Insights and Ideas

  • Water-Activated Stiffness: Using water temperature as the activation mechanism for stiffness change is a clever and safer alternative to electric heating, which can pose risks in surgical environments.
  • Bioplastic for Safety and Performance: FORMcard's biocompatibility, relatively low softening temperature, and tunable stiffness make it an excellent candidate for internal medical devices.
  • MEPCM for Thermal Safety: The MEPCM insulation is crucial for preventing thermal injury, effectively maintaining the manipulator's surface temperature near body temperature even when using hot water for activation. This significantly enhances the safety profile for in-vivo applications.
  • Integrated Design for NOTES: The design balances the need for a slim profile (essential for natural orifices) with the functional requirements of variable stiffness and tool delivery.

Key Results

The researchers fabricated a prototype and conducted several tests:

  • Mode Activation Time:
    • Flexible mode activation: 25 seconds.
  • Rigid mode activation: 20 seconds. These times are reported as shorter than some existing material-based variable stiffness endoscopes.

  • Bending Stiffness:

    • Flexible mode stiffness: 22.9–36.9 N·cm².
  • Rigid mode stiffness: 547.9–926.3 N·cm². This represents a significant improvement in stiffness in rigid mode.

  • Stiffness-Adjusting Ratio: 23.9–25.1 times. This indicates a substantial difference in stiffness between the two modes.

  • Surface Temperature Test: The MEPCM insulation kept the manipulator surface temperature below 41°C for approximately 18 minutes, well below the burn threshold of 44°C, providing a safe operating window.

Limitations and Future Work

  • Limitations:
    • The current study presents a conceptual design and prototype validation.
  • While FORMcard is biocompatible, extensive in-vivo testing and regulatory approval would be necessary for clinical use.
  • The paper mentions that the stiffness values in flexible and rigid modes differ from a suggested range for NOTES endoscopes, implying further optimization might be needed for specific applications.
  • Future Work:
    • Develop a fully functional endoscope with embedded cameras and other necessary electronics.
  • Test the device in realistic surgical scenarios (ex vivo and in vivo).
  • Proceed towards wider clinical applications.

Summary updated