A Noncontact ECG Sensing System With a Micropower, Ultrahigh Impedance Front-End, and BLE Connectivity
Abstract
Capacitive noncontact electrodes offer a promising solution to improve comfort and unobtrusiveness in continuous monitoring ECG systems which are becoming more widespread in wearable devices. Most implementations based on off-the-shelf components require rather large electrodes and power budgets while custom application-specific integrated circuits (ASICs) face significantly higher adoption challenges. In this article, we present a novel micropower ultrahigh impedance front-end with an equivalent input capacitance of 70 fF in the ECG band, enabling contactless ECG acquisition with electrodes as small as 78.5 mm2 with a power consumption of only <inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">$15 \, \mu \text{W}$ </tex-math></inline-formula> . The front-end implements a novel power supply bootstrapping architecture and is integrated in a full acquisition system which includes a 24-bit analog to digital converter (ADC) and a bluetooth low energy (BLE)-capable Cortex M4 processor, achieving up to ten days of continuous operation on a 600 mAh PR44 battery.
BibTeX
@article{guermandi2024noncontact,
author = {Marco Guermandi and Simone Benatti and Luca Benini},
title = {{A Noncontact ECG Sensing System With a Micropower, Ultrahigh Impedance Front-End, and BLE Connectivity}},
journal = {IEEE Sensors Journal},
year = {2024},
doi = {10.1109/jsen.2023.3347100},
}Ask Althea to put this paper to the test.