A NIM PET/CT phantom for evaluating the PET image quality of micro-lesions and the performance parameters of CT

Shujie Lu, Peng Zhang, Chengwei Li, Jie Sun, Wenli Liu, Pu ZhangBmc Medical Imaging 2021

TL;DR

This paper introduces a novel NIM PET/CT phantom designed to overcome the limitations of existing phantoms in evaluating the image quality of micro-lesions and CT performance. The new phantom allows for simultaneous PET and CT assessment, and its inclusion of smaller spheres (down to 4mm) enables more precise evaluation of micro-lesion detection capabilities across different PET/CT systems and reconstruction algorithms.

What Problem Does This Paper Address?

Positron Emission Tomography (PET) and Computed Tomography (CT) are crucial imaging modalities in medical diagnostics, particularly for tumor detection, staging, and treatment response evaluation. PET/CT systems combine these technologies to leverage their complementary strengths. However, accurately assessing the performance of these systems, especially their ability to detect small lesions (micro-lesions), is challenging. Existing phantoms, like the NEMA IEC Body phantom, have limitations: their smallest spheres are 10mm, making them unsuitable for evaluating micro-lesion detection. Furthermore, they are primarily designed for PET and offer limited CT performance evaluation. This gap hinders the precise comparison and optimization of different PET/CT systems and reconstruction techniques for early and accurate diagnosis.

Methods

The researchers developed a new PET/CT phantom, the NIM PET/CT phantom, which integrates both PET and CT imaging modules. The PET module is similar to the NEMA IEC Body phantom but includes additional smaller spheres with diameters of 4mm and 7mm, alongside the standard 10mm, 13mm, 17mm, 22mm, 28mm, and 37mm spheres. These spheres are filled with either 18F-FDG (hot lesions) or non-radioactive water (cold lesions) at a 4:1 sphere-to-background activity concentration. The CT module features low-contrast inserts with resolutions of 0.5%, 1.0%, and 1.5% to assess CT performance.

To evaluate the phantom's effectiveness, they compared its performance with the NEMA IEC Body phantom using three different PET/CT systems. They also investigated the impact of various PET image reconstruction algorithms (OSEM, OSEM-PSF, OSEM-TOF, and OSEM-PSF-TOF) on image quality metrics like contrast recovery coefficient (CRC) and background variability. Quantitative analysis involved defining regions of interest (ROIs) on both PET and CT images to measure these parameters.

Datasets

The study utilized a custom-designed NIM PET/CT phantom and the standard NEMA IEC Body phantom. These phantoms served as the primary "datasets" for evaluating PET/CT system performance. No external imaging datasets were used for training or evaluation.

Key Insights and Ideas

  • Micro-lesion detection is critical: The inclusion of 4mm and 7mm spheres in the NIM phantom is a significant advancement, directly addressing the need to evaluate the detection of clinically relevant micro-lesions, which are often missed by older phantoms.
  • Integrated PET/CT evaluation: The phantom's design allows for simultaneous assessment of both PET and CT performance, offering a more comprehensive quality control tool than separate phantoms.
  • Reconstruction algorithm impact: Different reconstruction algorithms have a substantial impact on image quality. Specifically, TOF (Time-of-Flight) and PSF (Point Spread Function) techniques, especially when combined (TOF-PSF), demonstrably improve image contrast, reduce noise, and enhance the detection of smaller structures.
  • System generalization: The NIM phantom's ability to be scanned on multiple PET/CT systems and yield comparable results highlights its potential for broad applicability in inter-system comparisons and quality assurance.

Key Results

  • The NIM PET/CT phantom successfully detected 7mm spheres, a capability not present in the NEMA IEC Body phantom.
  • Image quality (contrast and background variability) from the NIM phantom was generally consistent with the NEMA phantom for larger spheres.
  • The OSEM-PSF-TOF reconstruction algorithm showed the most significant improvements in image contrast and reduction of background variability, particularly for smaller spheres. For instance, the largest increase in image contrast for a 10mm sphere was 22.4% with TOF-PSF.
  • When tested across three different PET/CT systems, the NIM phantom demonstrated that system "C" generally provided superior image quality, including the detection of 4mm spheres, which were not detectable by systems "A" and "B".
  • The CT module of the NIM phantom effectively evaluated low-contrast resolutions of 1.5%, 1.0%, and 0.5%.

Limitations and Future Work

  • 4mm sphere detectability: While the NIM phantom includes a 4mm sphere, its detectability was not consistently achieved across all tested systems and reconstruction algorithms, indicating that further optimization might be needed for this smallest feature.
  • Shortcomings of the proposed phantom: The authors acknowledge that further research is needed to "eliminate those shortcomings and achieve more reliable outcomes," though specific shortcomings beyond the 4mm sphere detectability are not detailed.
  • Future work: The paper suggests that combining TOF with other reliable algorithms, along with CT-based attenuation and scatter correction, can further enhance PET/CT system selection and detailed medical imaging. The development of more advanced phantoms for comprehensive quality control of PET/CT systems is also implied.

Summary updated

BibTeX
@article{lu2021nim,
  author = {Shujie Lu and Peng Zhang and Chengwei Li and Jie Sun and Wenli Liu and Pu Zhang},
  title = {{A NIM PET/CT phantom for evaluating the PET image quality of micro-lesions and the performance parameters of CT}},
  journal = {Bmc Medical Imaging},
  year = {2021},
  doi = {10.1186/s12880-021-00683-4},
}