A new method for the calibration of the vertical scale of a stylus profilometer using multiple delta-layer films

Kyung-Joong Kim, Chil Seong Jung, Tae Eun HongMeasurement 2007 · DOI

Abstract

A new method for the vertical scale calibration in step height measurement has been established using a multiple delta-layer film. The vertical scale of a stylus profilometer can be traceable to the length unit by the certified film thickness in a multiple delta-layer. A Si/Ge multiple delta-layer film with six Si layers separated by Ge delta-layers was developed. The total thickness and delta-layer spacing were certified by high resolution transmission electron microscopy. Six craters with different depth were formed on the Si/Ge multiple delta-layer film by ion beam sputtering via SIMS depth profiling. The depths of the craters can be determined from the certified depth of the delta-layer and direct measurements using a stylus profilometer calibrated by a step-height standard. The vertical scale of a stylus profilometer can be calibrated by the slope and the offset of the measured crater depth fitted as a function of the certified nominal thickness.
BibTeX
@inproceedings{kim2007new,
  author = {Kyung-Joong Kim and Chil Seong Jung and Tae Eun Hong},
  title = {{A new method for the calibration of the vertical scale of a stylus profilometer using multiple delta-layer films}},
  booktitle = {Measurement},
  year = {2007},
  doi = {10.1088/0957-0233/18/9/002},
}

Loading related content...

Figure