A Near-Threshold, 0.16 nJ/b OOK-Transmitter With 0.18 nJ/b Noise-Cancelling Super-Regenerative Receiver for the Medical Implant Communications Service
Abstract
A 0.16 nJ/b MICS transmitter and 0.18 nJ/b super-regenerative receiver are demonstrated, where each is specifically designed to operate in the near-threshold region. The low-VDD transmitter utilizes a sub-harmonic injection-locked ring oscillator, edge combiner for frequency multiplication, and class-C power amplifier. The low-VDD receiver introduces a replica super-regenerative receiver as a method to reject common-mode noise sources, such as supply/substrate coupling, thereby reducing undesired self-oscillations and improving BER. Designed in a 90-nm CMOS process, the test-chip measurements show a sensitivity of -80 dBm at 500 kb/s and -65 dBm at 1 Mb/s, respectively, at a BER less than 10⁻³, with 340 μW total power.
BibTeX
@article{ma2013near,
author = {Chao Ma and Changhui Hu and Jiao Cheng and Lingli Xia and Patrick Yin Chiang},
title = {{A Near-Threshold, 0.16 nJ/b OOK-Transmitter With 0.18 nJ/b Noise-Cancelling Super-Regenerative Receiver for the Medical Implant Communications Service}},
journal = {IEEE Transactions on Biomedical Circuits and Systems},
year = {2013},
doi = {10.1109/tbcas.2013.2253555},
}Ask Althea to put this paper to the test.