TL;DR
This paper introduces a novel, high-resolution system for precisely measuring the temperature changes induced by micron-sized laser spots. This system overcomes previous limitations in evaluating laser-tissue interactions, providing crucial reference data for understanding photothermal effects in applications like optical nerve stimulation.
What Problem Does This Paper Address?
The precise control and understanding of laser-tissue interactions are critical for advancements in medical treatments and neural stimulation technologies, particularly in areas like cochlear implants. While the photothermal effect (heat generated by laser absorption) is known to play a role in nerve excitation, existing methods for evaluating this effect are often subjective, lack spatial resolution, or are incompatible with the micron-sized laser beams used in research. Specifically, measuring temperature changes from micron-sized laser spots on biological tissues is challenging due to limitations in sensor size, non-contact measurement techniques (like infrared), and the difficulty of precisely aligning micron-scale lasers with sensors. This paper aims to fill this gap by developing a system capable of quantitatively evaluating the photothermal effect of micron-sized lasers with high temporal and spatial resolution.
Methods
The core of the system is a micron-scale Pd/Cr thin-film thermocouple array fabricated on a freestanding Si₃N₄ platform. This design allows for sensitive temperature detection at the junction of the Pd and Cr films. To achieve precise laser targeting, the system incorporates a 3D positioning module with a movement accuracy of 0.03 mm. This module, coupled with real-time monitoring of the thermocouple's output voltage, allows for accurate alignment of the micron-sized laser beam with the thermocouple junction.
The data acquisition and processing are handled by a nanovoltmeter (Keithley 2182A) and a multiplexer, controlled by LabVIEW software. The multiplexer sequentially activates each thermocouple in the array, enabling the nanovoltmeter to measure the microvolt-level output voltage, which is then converted to temperature change via the Seebeck coefficient. The system can accommodate pulsed lasers with adjustable parameters (wavelength, pulse width, repetition frequency, and power) across six different wavelengths (450 nm to 1064 nm).
Datasets
Not discussed in the paper. The paper focuses on developing and validating a measurement system and generating reference data by directly irradiating the thermocouple sensors with lasers.
Key Insights and Ideas
- High-Resolution Photothermal Measurement: The development of a micron-scale thermocouple array combined with precise 3D laser positioning provides an unprecedented ability to measure photothermal effects at the micron scale.
- Wavelength-Dependent Photothermal Effects: The study reveals a complex relationship between laser wavelength and photothermal effect. For visible light (450-810 nm), shorter wavelengths generally lead to stronger photothermal effects. However, in the infrared range (980-1064 nm), longer wavelengths appear to be more effective. This finding is unexpected and warrants further investigation.
- Pulse Frequency vs. Intensity: For most wavelengths, laser intensity (peak power) has a more significant impact on local temperature rise than pulse frequency. However, for the 1064 nm laser, the opposite trend is observed, suggesting that the photothermal effect for this wavelength is more dependent on irradiation time.
- Stability of Light-to-Heat Conversion: The linear relationship between temperature change and energy density (or peak power) at a fixed pulse frequency indicates a stable ratio of photothermal conversion, making the system reliable for quantitative measurements.
Key Results
- The developed system can quantitatively evaluate the photothermal effect of micron-sized pulsed lasers with high temporal and spatial resolution.
- Experiments were conducted using lasers with wavelengths from 450 nm to 1064 nm and various pulse frequencies (≥50 Hz).
- The temperature rise of the Pd/Cr thermocouple was found to be linearly proportional to the peak power and average energy density of the laser when the pulse frequency was fixed.
- Table 1 provides specific data on the peak power required to achieve a 15 °C and 20 °C temperature increase for different wavelengths and pulse frequencies, serving as valuable reference data. For instance, to achieve a 15 °C increase at 500 Hz, the peak power required for 450 nm is 84.6 mW, while for 810 nm it's 981.5 mW.
Limitations and Future Work
- Lack of Biological Tissue Experiments: The current study focuses on laser-thermocouple interactions. While the authors infer that the system can be applied to tissue, direct experiments with biological tissues are not yet presented.
- Interference from Thermocouple Heating: The authors acknowledge that the thermocouple itself can heat up, potentially interfering with measurements of actual tissue temperature. They address this by designing very small (3 µm x 3 µm x 100 nm) thermocouple junctions to minimize this self-heating effect.
- Mechanism of Infrared Photothermal Effect: The "unusual high photothermal effect" of infrared lasers requires further investigation to understand the underlying mechanisms.
- Future Applications: The authors envision this system being used to establish mathematical models for photothermal effects in laser-tissue interactions and to quantitatively evaluate the role of photothermal effects in laser-neuron stimulation, advancing research in areas like optical cochlear implants.