A low-cost and accurate interface for four-electrode conductivity sensors
Abstract
This paper presents a low-cost and accurate interface for four-electrode conductivity sensors. The interface mainly consists of an analog front-end, a multiplexer, and a voltage-to-time converter. The analog front-end is used to provide a controlled excitation voltage for the sensor and to convert the sensor signal (conductance) into a voltage signal. The voltage-to-time period converter acts as an asynchronous converter for the sensor signals (voltage), which employs a relaxation oscillator and outputs a period-modulated signal. Experimental results over a conductance range of 0.1 /spl mu/S to 20 mS show a random error of 1.6/spl times/10/sup -5/ and a systematic error of 6.6/spl times/10/sup -5/ for a measurement time of 110 ms.
BibTeX
@article{li2005low,
author = {Xiujun Li and Gerard C. M. Meijer},
title = {{A low-cost and accurate interface for four-electrode conductivity sensors}},
journal = {IEEE Trans. Instrumentation \& Measurement},
year = {2005},
url = {https://openreview.net/forum?id=wgQB7HtNI1A},
}Ask Althea to put this paper to the test.