Abstract
Imaging is used to map activity across populations of neurons. Microscopes with cellular resolution have small (< 1 millimeter) fields of view and cannot simultaneously image activity distributed across multiple brain areas. Typical large field of view microscopes do not resolve single cells, especially in the axial dimension. We developed a 2-photon random access mesoscope (2p-RAM) that allows high-resolution imaging anywhere within a volume spanning multiple brain areas (Ø 5 mm × 1 mm cylinder). 2p-RAM resolution is near diffraction limited (lateral, 0.66 μm, axial 4.09 μm at the center; excitation wavelength = 970 nm; numerical aperture = 0.6) over a large range of excitation wavelengths. A fast three-dimensional scanning system allows efficient sampling of neural activity in arbitrary regions of interest across the entire imaging volume. We illustrate the use of the 2p-RAM by imaging neural activity in multiple, non-contiguous brain areas in transgenic mice expressing protein calcium sensors.
Introduction
In systems neuroscience, understanding how complex behaviors emerge requires monitoring the coordinated activity of neurons distributed across the brain. However, researchers have long faced a fundamental trade-off between scale and resolution. Standard two-photon (2P) microscopes offer the subcellular resolution necessary to distinguish individual synapses and neurons but are limited to a field of view (FOV) of less than 1 mm. Conversely, wide-field imaging techniques can capture the entire mouse cortex but lack the axial resolution to isolate individual neurons in scattering tissue.
To bridge this gap, the Svoboda lab at Janelia Research Campus developed the 2-photon random access mesoscope (2p-RAM). This instrument provides a 5 mm FOV—large enough to encompass multiple functional brain regions—while maintaining the diffraction-limited resolution required to resolve individual neurons and their processes.
Technical Innovations in the 2p-RAM
The 2p-RAM achieves its performance through several custom-engineered optical innovations designed to maintain high numerical aperture (NA) across a massive imaging volume.
Virtually Conjugated Galvo Pair (VCGP)
Standard scanning systems often suffer from aberrations at large scan angles because the laser beam shifts away from the center of the objective's entrance pupil. The 2p-RAM utilizes a three-mirror VCGP system that ensures the beam remains stationary on the pupil even at the edges of the 5 mm FOV. This design prevents the blurring and loss of signal typically associated with large-scale scanning.
Remote Focus (RF) Unit
Traditional microscopes change focus by moving the objective lens, a process limited by the lens's mass. The 2p-RAM employs a Remote Focus unit, which uses a lightweight mirror on a voice coil to adjust the focal depth upstream of the objective. This allows the system to "jump" between different depths in milliseconds, enabling near-simultaneous imaging of different cortical layers or non-contiguous brain regions.
Optimized Collection Optics
To maximize signal-to-noise ratios, the system was designed with a high collection efficiency. By using an oil-coupled condenser system, the authors increased the numerical aperture of the light detection path to 1.4, ensuring that as many emitted photons as possible reach the photomultiplier tubes (PMTs).
Performance and Validation
The 2p-RAM was benchmarked using both calibration targets and in vivo experiments in transgenic mice.
- Resolution: The system achieves a lateral resolution of 0.66 μm and an axial resolution of 4.09 μm at the center of the FOV. Crucially, this resolution remains sufficient to distinguish individual neurons even at the extreme edges of the 5 mm diameter cylinder.
- Imaging Volume: The total accessible volume (5 mm diameter x 1 mm depth) is approximately 100 times larger than that of a conventional 2P microscope. In mice expressing nuclear-localized GFP (H2B-GFP), the authors estimated that over 200,000 individual neurons are accessible within a single imaging volume.
- Functional Throughput: Using GCaMP6, the researchers demonstrated the ability to track the activity of over 3,000 neurons across the cortex at a 1.9 Hz frame rate. By utilizing the random-access capabilities to target smaller sub-regions, they achieved sampling rates of 9.6 Hz across four distinct, non-contiguous brain areas.
Discussion and Limitations
A key insight from this work is the efficiency of serial scanning. The authors argue that using a single high-power beam is more effective than multiplexing (splitting the beam), as the primary bottleneck in in vivo imaging is brain heating (limited to ~200 mW). Serial scanning ensures that the maximum allowable power is concentrated where it is needed most.
Despite its advantages, the 2p-RAM faces some constraints. The remote focus unit, while fast, cannot correct for field curvature within a single high-speed resonant scan line, leading to minor tilting at the FOV edges. Additionally, variations in PMT sensitivity across the detector surface require computational correction to ensure uniform brightness across the image.
Conclusion
The 2p-RAM represents a significant advancement in optical neuroimaging, providing the first tool capable of monitoring neural population dynamics at cellular resolution across the mesoscale. By enabling simultaneous access to multiple cortical areas, it allows researchers to investigate how information is integrated across the brain during complex behaviors.
Summary updated