A highly sensitive wide-range weak current detection circuit for implantable glucose monitoring

Xi Tan, Sizheng Chen, Xiao Yan, Yibo Fan, Hao Min, Junyu WangIEICE Electron. Express 2016 · DOI

Abstract

A highly sensitive weak current detection circuit for implantable glucose monitoring is presented. The circuit consists of a current splitter, an integrator, a sampling/holding (S/H) circuit, and an 8-bit successive approximation register (SAR) analog-to-digital convertor (ADC). T-switches, source voltage shifting and correlated double sampling (CDS) technology are adopted to reduce the effects of leakage currents, charge injection, dc offset and low-frequency 1/f noise. The work range is divided into 5 sub-ranges to reduce the complexity and power consumption of current detection. The chip was fabricated with 0.13 µm CMOS technology. Test results show that the circuit can detect weak signal from 10 fA to 1 nA with 1.5 V supply and 7.2 µW power consumption. The tested leakage current of the design is 4 fA.

Introduction

Continuous glucose monitoring (CGM) via wireless implantable sensors offers a significant improvement over traditional finger-prick tests, which only provide snapshots of blood glucose levels. However, these enzymatic sensors generate extremely weak "Faradic currents," often in the femtoampere (fA) to nanoampere (nA) range. Designing a readout circuit that is sensitive enough for fA signals yet robust enough for nA signals—all while maintaining ultra-low power consumption for implantation—is a critical engineering challenge.

System Architecture

The proposed system is designed to handle a massive detection range (10 fA to 1 nA) using a multi-range integration technique. The overall architecture, shown in Figure 1, includes a front-end for power and data modulation, and a baseband containing the weak current detection circuit.

Block diagram of the implantable monitoring system, highlighting the weak current detection circuit.

The detection pipeline consists of a current splitter for calibration, a switched-capacitor integrator, a sampling/holding (S/H) circuit, and an 8-bit successive approximation register (SAR) analog-to-digital converter (ADC). By dividing the work into five sub-ranges and adjusting the integration time, the system can "zoom in" on tiny currents or "zoom out" for larger ones without requiring a high-resolution, power-hungry ADC.

Circuit Design and Noise Reduction

To achieve high sensitivity, the circuit employs several specialized techniques to combat leakage and noise. A key component is the switched-capacitor integrator, which converts the input current ($I_{in}$) into a measurable voltage.

Schematic of the switched-capacitor integrator circuit using correlated double sampling (CDS) and specialized switching phases.

Key Innovations:

  • T-switches: At the femtoampere level, standard MOSFET switches leak significantly even when "off." The design utilizes a T-shaped switch configuration and source voltage shifting to reduce internal leakage to a measured 4 fA.
  • Correlated Double Sampling (CDS): This technique is integrated into the switching sequence (as seen in the $\Phi$ phases of Figure 2) to cancel DC offsets and low-frequency 1/f "flicker" noise.
  • Single-Ended Efficiency: To save power, the researchers opted for single-ended structures rather than fully differential ones, which consume significantly more energy.

Results and Discussion

Fabricated in 0.13 µm CMOS technology, the chip occupies a compact area of 1.2 × 1.8 mm². Key performance metrics include:

  • Detection Range: 10 fA to 1 nA (a 100,000x dynamic range).
  • Power Consumption: 7.2 µW at a 1.5 V supply, making it ideal for RFID-powered or battery-constrained implants.
  • ADC Accuracy: The 8-bit SAR ADC achieved an Effective Number of Bits (ENOB) of 7.8 bits, performing near its theoretical limit.

While the silicon characterization is successful, future work will focus on in vivo testing to evaluate performance in the corrosive biological environment of the human body and ensure long-term stability.

Summary updated

BibTeX
@article{tan2016highly,
  author = {Xi Tan and Sizheng Chen and Xiao Yan and Yibo Fan and Hao Min and Junyu Wang},
  title = {{A highly sensitive wide-range weak current detection circuit for implantable glucose monitoring}},
  journal = {IEICE Electron. Express},
  year = {2016},
  doi = {10.1587/elex.13.20150616},
}

Loading related content...

Figure