A flexible adhesive surface electrode array capable of cervical electroneurography during a sequential autonomic stress challenge

Yifeng Bu, Jonas Kurniawan, Jacob S. Prince, Andrew K. L. Nguyen, Brandon Ho, Nathan L. J. Sit, Timothy V. Pham, Vincent M. Wu, Boris Tjhia, Andrew Shin, Tsung‐Chin Wu, Xin Tu, Ramesh R. Rao, Todd P. Coleman, Imanuel LermanScientific 2022 · DOI

Abstract

This study introduces a flexible, adhesive-integrated electrode array developed to enable non-invasive monitoring of cervical nerve activity. The device utilizes silver-silver chloride (Ag/AgCl) electrodes combined with a customized biopotential data acquisition unit and graphical user interface (GUI) for real-time visualization. Preliminary testing demonstrated that this electrode design achieves a high signal-to-noise ratio during cervical neural recordings. To demonstrate the capability of the surface electrodes to detect changes in cervical neuronal activity, the cold-pressor test (CPT) and a timed respiratory challenge were employed as autonomic stressors. This recording technique, termed Cervical Electroneurography (CEN), utilized a custom spike-sorting algorithm to classify neural activity. Results showed rostral-to-caudal channel position-specific firing patterns and biotype-specific changes in average CEN firing. This work paves the way for ambulatory CEN recording devices to monitor autonomic dysregulation in clinical and healthy populations.

Introduction

The Autonomic Nervous System (ANS) acts as the body's internal autopilot, regulating essential functions such as heart rate, digestion, and respiratory rhythm. Dysregulation of the ANS is a significant factor in various conditions, including PTSD, diabetic neuropathy, and rheumatoid arthritis. Traditionally, monitoring the ANS has relied on indirect measures like heart rate variability (HRV) or invasive surgical procedures to implant "cuff" electrodes.

This research addresses the need for a non-invasive, high-resolution method to "listen" to cervical nerves—specifically the vagus nerve and sympathetic chain—during physiological stress. The resulting technique, Cervical Electroneurography (CEN), uses a skin-conformal electrode array to capture direct neural signals through the skin.

Device Design and Fabrication

The electrode array was engineered using a hybrid fabrication approach to ensure both high signal fidelity and mechanical flexibility. The device features thin-film microfabricated gold (Cr/Au) interconnects for high-resolution routing and thick-film screen-printed silver-silver chloride (Ag/AgCl) electrodes to maintain low impedance.

Exploded view and fabrication stages of the flexible electrode array.

The array is embedded in a "peel-and-stick" silicone substrate (Ecoflex/Silbione) that conforms to the curves of the neck. This allows for robust adhesion and signal stability even during head rotation and flexion. The system maintained a skin-to-electrode impedance of ≤ 1.6 kΩ over a two-hour period, which is optimal for high-quality physiological recording.

Experimental Methodology

The study involved 10 healthy participants who underwent two specific stress protocols while being monitored by the CEN array, ECG, and a respiration belt.

Experimental setup showing the placement of the CEN array on the neck and other physiological sensors.
  1. Cold Pressor Test (CPT): Subjects immersed their hand in ice water (3–7°C) to trigger a sympathetic "fight or flight" response.
  2. Timed Respiratory Challenge: A controlled breathing exercise (5s inhale, 5s exhale) was used to modulate the parasympathetic system.

Signals were captured at 8 kHz and processed through a custom spike-sorting pipeline involving bandpass filtering, t-SNE dimensionality reduction, and unsupervised clustering (DBSCAN) to separate neural spikes from background noise and muscle activity (EMG).

Key Insights and Results

  • Spatiotemporal Mapping: The array's four channels, arranged from the top (rostral) to the bottom (caudal) of the neck, revealed position-specific firing patterns. Caudal channels reacted immediately to the ice water challenge, while rostral channels showed a delayed "rebound" response.
  • Autonomic Biotypes: The study identified two distinct groups: "Increasers" (CPTi) and "Decreasers" (CPTd), based on heart rate reactions to pain. CEN firing patterns matched these biotypes; CPTi subjects showed increased nerve firing during breathing challenges, while CPTd subjects showed a decrease.
  • Surface Fidelity: A major finding was that surface electrodes, when designed correctly, can capture enough detail to perform spike sorting—a task typically requiring invasive probes. All channels showed a significant increase in neural firing frequency (p < 0.05) during the initial phase of the cold water challenge.

Conclusion and Future Work

This pilot study demonstrates the feasibility of using flexible, adhesive surface electrode arrays for non-invasive cervical electroneurography. While the current sample size is small, the results suggest that CEN can identify unique neural "fingerprints" related to autonomic stress responses. Future work will focus on developing wireless, ambulatory CEN devices for long-term monitoring, which could provide early notification of autonomic dysregulation in clinical disease states.

Reference: Bu, Y., Kurniawan, J.F., Prince, J. et al. A flexible adhesive surface electrode array capable of cervical electroneurography during a sequential autonomic stress challenge. Sci Rep 12, 19467 (2022).

Summary updated

Abstract

This study introduces a flexible, adhesive-integrated electrode array that was developed to enable non-invasive monitoring of cervical nerve activity. The device uses silver-silver chloride as the electrode material of choice and combines it with an electrode array consisting of a customized biopotential data acquisition unit and integrated graphical user interface (GUI) for visualization of real-time monitoring. Preliminary testing demonstrated this electrode design can achieve a high signal to noise ratio during cervical neural recordings. To demonstrate the capability of the surface electrodes to detect changes in cervical neuronal activity, the cold-pressor test (CPT) and a timed respiratory challenge were employed as stressors to the autonomic nervous system. This sensor system recording, a new technique, was termed Cervical Electroneurography (CEN). By applying a custom spike sorting algorithm to the electrode measurements, neural activity was classified in two ways: (1) pre-to-post CPT, and (2) during a timed respiratory challenge. Unique to this work: (1) rostral to caudal channel position-specific (cephalad to caudal) firing patterns and (2) cross challenge biotype-specific change in average CEN firing, were observed with both CPT and the timed respiratory challenge. Future work is planned to develop an ambulatory CEN recording device that could provide immediate notification of autonomic nervous system activity changes that might indicate autonomic dysregulation in healthy subjects and clinical disease states.
BibTeX
@inproceedings{bu2022flexible,
  author = {Yifeng Bu and Jonas Kurniawan and Jacob S. Prince and Andrew K. L. Nguyen and Brandon Ho and Nathan L. J. Sit and Timothy V. Pham and Vincent M. Wu and Boris Tjhia and Andrew Shin and Tsung‐Chin Wu and Xin Tu and Ramesh R. Rao and Todd P. Coleman and Imanuel Lerman},
  title = {{A flexible adhesive surface electrode array capable of cervical electroneurography during a sequential autonomic stress challenge}},
  booktitle = {Scientific},
  year = {2022},
  doi = {10.1038/s41598-022-21817-w},
}
Figure