A combined application of Deep Learning Recon and MRI mute technique
Abstract
Motivation: MRI mute technique reduces scanning noise by lowering the gradient slew rate which increase echo spacing, resulting in image blurring and longer scan time. Goal(s): To design a scanning method that simultaneously reduces scan time and scanning noise without compromising image SNR and clarity. Approach: Develop a DL-Recon post-processing model to enhance SNR and clarity of images which are acquired from optimized knee scanning protocol with mute technique. Results: The proposed scanning method improves SNR about 38% and reduces scan time and noise sound pressure separately about 44.8% and 86%. Impact: Combining DL-Recon with mute sequences help doctors to diagnose more patients. It also enhances the success of clinical scans by decreasing scan noise to improve patient comfort. This is a successful application of DL-Recon with mute scanning technique.
BibTeX
@inproceedings{wang2024combined,
author = {Hongbin Wang and Weinan Tang and Jianghua Wu and Wei Xi},
title = {{A combined application of Deep Learning Recon and MRI mute technique}},
booktitle = {Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition},
year = {2024},
doi = {10.58530/2024/4660},