A Bi-Polar Current Source With High Short-Term Stability for Tsinghua Tabletop Kibble Balance

Kang Ma, Xiaohu Liu, Wei Zhao, Songling Huang, Shisong LiIEEE Trans. Instrumentation & Measurement 2024 · DOI

TL;DR

This paper introduces a novel, digitally-controlled, two-stage current source designed for high-precision Kibble balance experiments. By combining a large-range, low-resolution current source with a small-range, high-resolution one, the system achieves nA/A stability, crucial for accurate mass metrology. This approach offers a practical and cost-effective alternative to custom-built, high-precision current sources.

What Problem Does This Paper Address?

The Kibble balance is a cornerstone of modern mass metrology, serving as the primary method for realizing the kilogram based on fundamental physical constants. Achieving the required measurement uncertainty (around 10⁻⁹) necessitates extremely stable and precise electrical quantities, including the current used in the weighing phase. While many Kibble balance experiments rely on custom-designed current sources, these are often complex and time-consuming to build. Existing commercial current sources, even high-end models, typically fall short of the stringent stability and resolution requirements for this application. This paper addresses the need for a readily achievable, high-performance current source that can meet these demands without requiring extensive custom hardware development.

Methods

The core innovation is a digital-feedback two-stage current source (DTCS). This system cleverly combines two commercially available current sources:

  1. Main Current Source (MCS): A larger-range, lower-resolution source that provides the bulk of the current (e.g., up to 20 mA).
  2. Compensating Current Source (CCS): A smaller-range, much higher-resolution source that acts as a fine-tuner.

The two sources are connected in parallel to supply current to a stable resistor and the Kibble balance coil. The system operates in two primary modes:

  • Constant Force (CF) Scheme: Used during the weighing phase where the current is adjusted to maintain a constant force. Here, the DTCS compensates for mechanical drifts.
  • Constant Current (CC) Scheme: Used when the current needs to be precisely stable, particularly during velocity measurements to avoid induced voltage errors. In this mode, a digital voltmeter (DVM) measures the voltage across a calibrated standard resistor. The difference between this measured current and the desired setpoint is fed into a Proportional-Integral-Derivative (PID) controller. The PID controller then adjusts the output of the CCS to precisely compensate for any fluctuations or noise from the MCS.

The entire feedback loop, including the DVM and PID controller, is implemented digitally using LabVIEW. This allows for precise control and real-time adjustment of the current.

Datasets

Not discussed in the paper. The paper focuses on experimental validation of the developed current source using its own output and calibrated measurement equipment.

Key Insights and Ideas

  • Leveraging Commercial Off-the-Shelf (COTS) Components: The paper demonstrates that by intelligently combining COTS current sources with a digital feedback loop, it's possible to achieve performance levels typically requiring custom-built systems. This significantly lowers the barrier to entry for high-precision metrology.
  • Digital Feedback for Stability: The use of a digital PID controller to actively compensate for the noise and drift of a primary current source is a key enabler. This allows the system to achieve stability far beyond what either source could achieve individually.
  • Two-Stage Compensation Strategy: The separation of bulk current provision (MCS) and fine-tuning (CCS) is an elegant solution. The CCS, with its high resolution, can effectively cancel out the low-frequency noise (1/f noise) and short-term drifts of the MCS.
  • Notch Filtering via Averaging: The paper highlights how a moving-average filter in the feedback loop can introduce deliberate notches in the frequency spectrum of the output current. While this can suppress specific frequencies, it also demonstrates the trade-offs in filter design and their impact on the noise spectrum.
  • CC Scheme is More Demanding: The analysis clearly shows that the Constant Current (CC) scheme, crucial for minimizing systematic errors in velocity measurements, imposes more stringent stability requirements on the current source than the Constant Force (CF) scheme.

Key Results

  • The developed DTCS achieves a short-term stability of 20 nA/A at an integration time of 100 seconds.
  • With longer integration times (around 30 minutes), the stability can reach 1 nA/A.
  • The system effectively suppresses the 1/f noise of the main current source (MCS), resulting in a significantly lower noise spectral density (NSD) at low frequencies.
  • The DTCS can operate with output currents ranging from several mA up to 20 mA, meeting the typical requirements for Kibble balance experiments.
  • Experimental tests confirm the DTCS's suitability for both Constant Force (CF) and Constant Current (CC) operation schemes.

Limitations and Future Work

  • DVM Precision: The paper notes that the precision of the DVM is currently a major limitation for further improving the proposed DTCS.
  • PJVS Integration: The Programmable Josephson Quantum Voltage Standard (PJVS) system has not yet been integrated, which could further enhance the accuracy and stability of the feedback loop.
  • Loop Time Reduction: Accelerating the residual voltage measurement and improving the DVM's update rate are identified as areas for reducing the servo control loop time, which would further enhance stability.
  • PID Parameter Optimization: While optimal PID parameters were found for the tested conditions, further experimental investigation might be needed for different operating points or environmental conditions.
  • Notch Filter Design: The paper acknowledges that the placement and depth of notches in the frequency spectrum, introduced by the moving-average filter, can be adjusted by modifying PID parameters and filter settings, suggesting further exploration in this area.

Summary updated

BibTeX
@article{ma2024bi,
  author = {Kang Ma and Xiaohu Liu and Wei Zhao and Songling Huang and Shisong Li},
  title = {{A Bi-Polar Current Source With High Short-Term Stability for Tsinghua Tabletop Kibble Balance}},
  journal = {IEEE Trans. Instrumentation \& Measurement},
  year = {2024},
  doi = {10.1109/tim.2024.3481546},
}