A 60-channel high-density flexible receive array for pediatric abdominal MRI
Abstract
Motivation: Conventional MRI coils offer suboptimal parallel imaging performance for young children. Goal(s): Our goal was to enhance imaging acceleration by dedicated flexible high-density coil design for pediatric patients at 3T. Approach: We design, construct, and evaluate a highly flexible small element dense array constituted by dual turn loops with minimum inter-component interference layout design. Results: Both phantom and in-vivo studies demonstrated superior parallel imaging performance using the proposed coil. Impact: A dedicated high-channel count coil that allows highly localized coil sensitivities by minimum interference layout design may benefit small pediatric patients.
BibTeX
@inproceedings{lee2024channel,
author = {Wonje Lee and Yun-Jeong Stickle and Clyve Follante and Thomas Grafendorfer and Taeyoung Yang and Fraser Robb and Fan Zhang and Greig Scott and John M. Pauly and Shreyas Vasanawala and Ali Syed},
title = {{A 60-channel high-density flexible receive array for pediatric abdominal MRI}},
booktitle = {Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition},
year = {2024},
doi = {10.58530/2024/1035},
}Ask Althea to put this paper to the test.