A 10-Bit 300 kS/s Reference-Voltage Regulator Free SAR ADC for Wireless-Powered Implantable Medical Devices

Yongkui Yang, Jun Zhou, Xin Liu, Wang Ling GohSensors 2018 · DOI

TL;DR

This paper introduces a novel reference-voltage regulator-free SAR ADC designed for low-power, wireless-powered implantable medical devices. By employing a self-timed pre-charging technique, it eliminates the need for power-hungry regulators and bulky decoupling capacitors, significantly reducing power consumption and chip area while maintaining good performance under supply voltage fluctuations.

What Problem Does This Paper Address?

Implantable medical devices (IMDs) are increasingly reliant on wireless power transfer, which, even with advanced rectifiers, results in a fluctuating DC supply voltage. Traditional Successive Approximation Register (SAR) Analog-to-Digital Converters (ADCs) used in these devices require a stable reference voltage to operate accurately. This stability is typically achieved using a dedicated voltage regulator and decoupling capacitors. However, these components are power-hungry and occupy valuable chip area, which is a critical constraint for miniaturized IMDs. The power consumed by these regulators can account for a substantial portion (over 30%) of the total ADC power, and the decoupling capacitors increase silicon area and cost. This paper aims to overcome these limitations by designing a SAR ADC that can operate reliably without a dedicated reference-voltage regulator, thereby reducing power consumption and chip footprint.

Methods

The core innovation of this work is the reference-voltage regulator-free SAR ADC with self-timed pre-charging. Instead of relying on an external or internal regulator to provide a stable reference voltage, the proposed ADC utilizes a self-timed pre-charging mechanism. This technique dynamically charges the Digital-to-Analog Converter (DAC) capacitors using a current source until a charge detector signals that a pre-set charge level (Qset) has been reached. This pre-charging is controlled by a charge detector that monitors when the capacitor voltage crosses the reference voltage.

To mitigate the impact of supply voltage fluctuations on the pre-charging process, a two-phase pre-charging scheme is implemented. The first phase uses a large charging current for rapid charging, while the second phase employs a smaller current for fine-tuning. This approach aims to reduce charge variation caused by supply ripple and detector delay, ensuring that the differential charge variation remains below the Least Significant Bit (LSB) charge.

The ADC itself employs an asynchronous SAR logic with a charge-sharing DAC. The reference voltage is generated using a simple 2-transistor (2T) voltage reference circuit, which is less power-hungry than traditional regulators.

Datasets

Not discussed in the paper. The evaluation is based on chip fabrication and measurement results.

Key Insights and Ideas

  • Eliminating the Regulator is Key: The most significant insight is that by carefully designing the pre-charging and sampling stages, a SAR ADC can tolerate supply voltage fluctuations without a dedicated regulator, drastically improving power efficiency and reducing area.
  • Self-Timed Pre-Charging for Dynamic Control: The self-timed pre-charging mechanism is crucial. It allows the ADC to adapt its charging duration based on the actual supply conditions, rather than relying on a fixed, potentially over-engineered, regulator.
  • Two-Phase Pre-Charging for Robustness: The two-phase pre-charging strategy is a clever way to balance charging speed and accuracy. It addresses the trade-off between fast charging (needed for high sampling rates) and minimizing charge variation due to supply ripple.
  • Area and Power Savings are Synergistic: Removing the regulator and decoupling capacitor not only saves power directly but also reduces the overall silicon area, which is a critical factor for implantable devices.

Key Results

  • SNDR: The proposed SAR ADC achieves a Signal-to-Noise and Distortion Ratio (SNDR) of 53.32 dB operating at 0.8 V with a supply voltage fluctuation of 50 mVpp.
  • Power Consumption: The total power consumption, including the self-timed pre-charging circuits, is 2.72 µW at a sampling rate of 300 kS/s.
  • Figure of Merit (FOM): The total FOM is 23.9 fJ/conversion-step, which is competitive.
  • Area: The total area occupied by the proposed SAR ADC (Core1) is 0.105 mm².
  • Supply Voltage Fluctuation Tolerance: The design demonstrates good performance with a supply voltage fluctuation of 50 mVpp, achieving an ENOB of 8.56 bits, compared to 5.61 bits for a conventional ADC under the same conditions.
  • Power Savings: The self-timed pre-charging circuits consume only about 0.2 µW, which is significantly less than the typical 30% or more power consumption attributed to reference-voltage regulators in other designs.

Limitations and Future Work

  • Calibration for Gain Error: The paper acknowledges that the common part of charge variation (ACMQ) can cause a gain error, which needs to be removed by calibration. While this is standard practice, it adds complexity.
  • Sensitivity to Supply Ripple Frequency: While the paper demonstrates performance under sinusoidal fluctuations at 134.2 kHz and 13.56 MHz, the behavior at very high ripple frequencies or different ripple shapes is not extensively explored.
  • Future Work: The authors suggest that further optimization of the self-timed pre-charging circuits and the charge detector could lead to even better performance. The design is deemed adequate for medical devices, implying that further improvements could target higher resolution or sampling rates for more demanding applications.