A 1.8 m Class Pathfinder Raman LIDAR for the Northern Site of the Cherenkov Telescope Array Observatory—Technical Design

O. Ballester, O. Blanch, J. Boix, P. Calisse, A. Campoy-Ordaz, S. M. Colak, Vania Da Deppo, M. Doro, L. Font, Eudald Font-Pladevall, R. A. García, M. Gaug, R. Grau, Darko Kolar, A. López-Oramas, C. Maggio, Manel Martinez, O. Martı́nez, Victor Riu-Molinero, David Roman, S. Stanič, Júlia Tartera-Barberà, Santiago Ubach, Marko Zavrtanik, Miha ŽivecRemote 2025 · DOI

TL;DR

This paper details the technical design of the "pathfinder Barcelona Raman LIDAR" (pBRL), a crucial instrument for the Cherenkov Telescope Array Observatory (CTAO). The pBRL uses a 1.8m telescope and a pulsed laser to continuously monitor atmospheric conditions, which is essential for accurately interpreting high-energy gamma-ray observations.

What Problem Does This Paper Address?

High-energy gamma-ray astrophysics relies on detecting faint signals from cosmic events. These signals interact with Earth's atmosphere, creating showers of particles that emit Cherenkov light. The accuracy of reconstructing the energy and flux of these gamma rays is heavily dependent on understanding the atmospheric conditions, particularly aerosol and cloud scattering and extinction. Existing methods have limitations, and the CTAO, a next-generation observatory, requires a robust system for real-time atmospheric characterization. This paper addresses the need for a precise, steerable LIDAR system capable of profiling the atmosphere from the ground up to the lower stratosphere, specifically for the CTAO's northern site.

Methods

The pBRL is a Raman LIDAR system built upon a refurbished Cherenkov Light Ultraviolet Experiment (CLUE) telescope. Key technical choices include:

  • Coaxial Design: The laser beam is aligned with the telescope's optical axis to maximize overlap and improve near-range sensitivity.
  • 1.8m Telescope: A large parabolic mirror (f/1) is used for efficient light collection.
  • Pulsed Nd:YAG Laser: Emitting at 355 nm and 532 nm (frequency-doubled and tripled), these wavelengths are chosen to be within the Cherenkov light spectrum and avoid interference with optical telescopes.
  • Raman Channels: In addition to elastic scattering, Raman channels are used to measure aerosol scattering and extinction more accurately, allowing for the retrieval of the Ångström exponent.
  • Liquid Light Guide (LLG): An 8mm diameter LLG transports light from the telescope's focal plane to the polychromator.
  • Custom Polychromator: A sophisticated optical bench separates the light into four channels (two elastic, two Raman) using dichroic mirrors and interference filters.
  • Photomultiplier Tubes (PMTs): Used for photon detection, chosen for their availability and compatibility with existing IACT systems.
  • Near-Range System: A dedicated mini-telescope is incorporated to improve sensitivity at very close distances (below 500m).
  • Semi-Remote Operation: The system is controlled via a command-line interface and monitored using GRAFANA, allowing for efficient operation at the remote observatory site.

Datasets

Not discussed in the paper. The paper focuses on the technical design and calibration of the LIDAR instrument itself.

Key Insights and Ideas

  • Refurbishment as a Viable Path: Re-purposing existing astronomical equipment (like the CLUE telescope) can be a cost-effective and efficient way to develop specialized instruments.
  • Coaxial LIDAR Advantages: For atmospheric characterization relevant to IACTs, a coaxial design offers better near-range overlap and sensitivity compared to biaxial systems.
  • Importance of Raman Channels: While elastic LIDAR provides information, Raman channels are crucial for accurately discriminating between different aerosol types and achieving the required precision for atmospheric correction.
  • Balancing Performance and Practicality: The design choices (e.g., using N-BK7 glass, standard PMTs) reflect a balance between achieving high performance and managing costs and procurement complexity.
  • Long-Term Stability is Key: The instrument's design and testing emphasize stability over extended periods, crucial for continuous atmospheric monitoring.

Key Results

  • The pBRL was successfully designed, built, and tested at the Roque de los Muchachos Observatory (ORM) for a one-year period.
  • The instrument demonstrated stable operation and the ability to perform semi-remote controlled measurements.
  • Light leakage tests showed acceptable levels for Raman channels, crucial for accurate measurements.
  • The LLG transmission showed a slight degradation over 10 years, which is considered acceptable within the projected 15-year lifetime of the LIDAR.
  • The system's pointing and alignment procedures were developed and tested, showing stability over several months.

Limitations and Future Work

  • PMT Limitations: The current PMTs have a gain at the lower limit of the LOTR's photon-counting threshold and may need replacement with higher-gain PMTs adapted for specific wavelengths in a final version.
  • LLG Degradation: While acceptable, the LLG's transmission degradation over time suggests it may need replacement during the LIDAR's projected lifetime, requiring a spare to be provided.
  • Laser Power and Divergence: The current laser power and divergence might lead to saturation at short ranges, suggesting a future version could benefit from a laser with slightly smaller power and higher repetition rate, or potentially PMTs with gating capabilities.
  • Alignment Tediousness: The laser arm alignment procedure is described as tedious and could be improved with a steerable second guide mirror using piezoelectric motors.
  • Data Analysis: A detailed description of the data analysis carried out on the test data is outside the scope of this paper and will be presented in a follow-up article.

Summary updated

Abstract

This paper presents the technical design of the pathfinder Barcelona Raman LIDAR (pBRL) for the northern site of the Cherenkov Telescope Array Observatory (CTAO-N) located at the Roque de los Muchachos Observatory (ORM). The pBRL is developed for continuous atmospheric characterization, essential for correcting high-energy gamma-ray observations captured by Imaging Atmospheric Cherenkov Telescopes (IACTs). The LIDAR consists of a steerable telescope with a 1.8 m parabolic mirror and a pulsed Nd:YAG laser with frequency doubling and tripling. It emits at wavelengths of 355 nm and 532 nm to measure aerosol scattering and extinction through two elastic and Raman channels. Built upon a former Cherenkov Light Ultraviolet Experiment (CLUE) telescope, the pBRL’s design includes a Newtonian mirror configuration, a coaxial laser beam, a near-range system, a liquid light guide and a custom-made polychromator. During a one-year test at the ORM, the stability of the LIDAR and semi-remote-controlled operations were tested. This pathfinder leads the way to designing a final version of a CTAO Raman LIDAR which will provide real-time atmospheric monitoring and, as such, ensure the necessary accuracy of scientific data collected by the CTAO-N telescope array.
BibTeX
@inproceedings{ballester2025m,
  author = {O. Ballester and O. Blanch and J. Boix and P. Calisse and A. Campoy-Ordaz and S. M. Colak and Vania Da Deppo and M. Doro and L. Font and Eudald Font-Pladevall and R. A. García and M. Gaug and R. Grau and Darko Kolar and A. López-Oramas and C. Maggio and Manel Martinez and O. Martı́nez and Victor Riu-Molinero and David Roman and S. Stanič and Júlia Tartera-Barberà and Santiago Ubach and Marko Zavrtanik and Miha Živec},