Abstract
This study explores a wearable approach to quantifying physical load and muscle fatigue using a dual-sensor system. By integrating skin temperature sensing with electromyography (EMG), researchers aim to provide objective physiological markers for "katakori" (chronic shoulder stiffness). Experimental results from a repetitive exercise task demonstrate a clear correlation between increased muscle recruitment (iEMG) and rising skin temperature, suggesting a viable path for home-based muscle health monitoring.
Introduction
Chronic shoulder stiffness and lower back pain affect nearly 90% of the Japanese population. Despite its prevalence, "katakori" remains difficult to quantify objectively. While surface electromyography (EMG) is the standard for measuring muscle activity, it often fails to distinguish between active muscle use and cumulative fatigue. This research proposes a multi-modal approach, adding skin temperature monitoring to capture the thermal byproducts of muscle metabolism and blood flow changes.
Methodology
The prototype wearable system consists of two primary sensing modalities:
- Temperature Sensing: A K-type thermocouple paired with an AD595 compensation IC, calibrated to a precision of ±0.1°C.
- Electromyography (EMG): Surface electrodes connected to an instrumentation amplifier (INA114) with a 64dB gain. A PSoC (Programmable System on Chip) was used to implement a 120–320Hz band-pass filter to minimize motion artifacts.
Six healthy male subjects (ages 22–29) participated in a fatigue task involving 100 repetitions of vertical arm lifts with 1kg weights. Sensors were placed on the trapezius area, with the temperature sensor positioned between the EMG electrodes.
Results and Discussion
The system successfully tracked physiological changes across the rest, exercise, and recovery phases.
iEMG Trends
Integrated EMG (iEMG) values typically showed a slight initial decrease as subjects settled into a rhythmic movement, followed by a steady climb. This upward trend indicates the recruitment of additional motor units to compensate for fatiguing muscle fibers. Notably, higher iEMG values were often observed in the dominant shoulder.
Thermal Response
Skin temperature rose significantly in the majority of subjects during exercise. Unlike the electrical signal, temperature often remained elevated or continued to fluctuate during the recovery phase. This "thermal lag" suggests that heat dissipation provides a longer-lasting signature of physical load than EMG alone.
Conclusion
The study demonstrates that a low-power, wearable-scale system can capture high-fidelity physiological data comparable to laboratory equipment. Future work will focus on establishing definitive fatigue thresholds and testing the system on older populations or individuals with pre-existing chronic stiffness to refine its clinical utility.
Summary updated