NeurIPS 2026Coming to NeurIPS?Explore papers, orals & directions

3D fluorescence emission difference microscopy based on spatial light modulator

Guangyuan Zhao, Zihao Rong, Cuifang Kuang, Cheng Zheng, Xü LiuJournal of Innovative Optical Health Sciences 2016

Abstract

We report three-dimensional fluorescence emission difference (3D-FED) microscopy using a spatial light modulator (SLM). Zero phase, 0–2[Formula: see text] vortex phase and binary 0-pi phase are loaded on the SLM to generate the corresponding solid, doughnut and z-axis hollow excitation spot, respectively. Our technique achieves super-resolved image by subtracting three differently acquired images with proper subtractive factors. Detailed theoretical analysis and simulation tests are proceeded to testify the performance of 3D-FED. Also, the improvement of lateral and axial resolution is demonstrated by imaging 100[Formula: see text]nm fluorescent beads. The experiment yields lateral resolution of 140[Formula: see text]nm and axial resolution of approximate 380[Formula: see text]nm.
BibTeX
@article{zhao2016d,
  author = {Guangyuan Zhao and Zihao Rong and Cuifang Kuang and Cheng Zheng and Xü Liu},
  title = {{3D fluorescence emission difference microscopy based on spatial light modulator}},
  journal = {Journal of Innovative Optical Health Sciences},
  year = {2016},
  doi = {10.1142/s1793545816410030},
}

Loading related content...