33rd International Meeting of the German Society for Cardiovascular Engineering

Bert Arnrich, A. Albert, Johanna Walter, Frank Schön, M Rothe, Jürgen Ennker, Amir Khosravi, Andreas Liebold, M.s. Joharchi, K. David Wegner, Bernd Westphal, Gustav SteinhoffJournal of ExtraCorporeal Technology 2004 · DOI

Official logo of the Deutsche Gesellschaft für Kardiotechnik E. V.

Abstract

Blood cell trauma in cardiac surgery is primarily attributed to the use of the heart-lung machine and the oxygenator. Optimization of oxygenator design and membrane technology aims to reduce these alterations. This collection of research from the 33rd International Meeting of the German Society for Cardiovascular Engineering evaluates the impact of different oxygenator types and surgical techniques on blood cells and organ function markers. Analysis of a clinical database of 4,847 patients undergoing coronary artery bypass grafting (CABG) reveals significant differences in blood cell alterations across oxygenator groups. Furthermore, the separation of cardiotomy suction blood is shown to be a critical factor in reducing systemic inflammation and preserving platelet counts.

Introduction

Cardiopulmonary bypass (CPB) is a cornerstone of modern cardiac surgery, yet it remains inherently traumatic. When blood leaves the body and contacts synthetic surfaces, it triggers a massive inflammatory response, destroys blood cells (hemolysis), and activates the coagulation cascade. The researchers in this proceedings collection tackle several facets of this problem, ranging from biocompatibility engineering to surgical technique and advanced training simulations.

Methods and Datasets

The abstracts employ a variety of methodologies to address extracorporeal circulation (ECC) challenges:

  • Clinical Database Analysis: Data from 4,847 patients were analyzed to compare 13 oxygenator types, using a novel formula to normalize blood alterations per "transfer cycle" across the machine.
  • Prospective Randomized Trials: Studies compared standard CPB versus protocols where cardiotomy suction blood was separated and processed differently.
  • Hardware-in-the-Loop Simulation: Development of the "CardioSim" and "GREENBOX" physical simulators to mimic human hemodynamics for high-fidelity perfusionist training.
  • Ex Vivo Organ Models: Use of a modified Langendorff perfusion system for 24-hour preservation of isolated pig hearts.

Key Insights

The Surface Material Trade-off

There is no "perfect" surface. Researchers found that oxygenator modifications that best preserved platelets (PLT) often resulted in the highest white blood cell (WBC) activation, suggesting a fundamental trade-off in biocompatibility engineering.

Suction as an Inflammatory Driver

Cardiotomy suction (returning blood from the chest cavity to the pump) is a primary source of blood activation. Separating this blood significantly reduced systemic markers of inflammation like IL-6 (14 vs. 21 pg/mL) and improved platelet preservation (232 vs. 158 x10³/µL) compared to immediate retransfusion.

The "Aprotinin Mask"

A study on surface-modifying additives (SMA) found no clinical benefit in reducing blood loss. The authors suggest that the standard use of aprotinin (a drug used to reduce bleeding) might be so effective at suppressing inflammatory pathways that it "masks" any incremental benefit provided by expensive biocompatible coatings.

Key Results

  • Off-Pump Benefits: Off-pump surgery (OPCAB) showed significantly lower markers of myocardial ischemia (CK-MB, TropT) compared to on-pump surgery, though markers of contact activation remained similar.
  • Organ Preservation: The modified Langendorff system successfully maintained pig hearts in metabolic equilibrium for 24 hours, a significant jump from the standard 4-hour clinical ischemic window.
  • Clinical Outcomes: In a randomized trial of 150 patients, surface-modified circuits showed no significant difference in 24h blood loss (387 mL vs 399 mL) or total hospital stay compared to standard circuits.

Conclusion

Cardiotomy suction blood separation has a clear effect on reducing blood activation and preventing its negative systemic effects. While engineering improvements in oxygenators show promise in biochemical markers, their clinical impact is often obscured by pharmacological interventions. Further studies are needed to determine if retransfusion of cardiotomy suction blood without cell saving is indicated in routine CABG.

Summary updated

Abstract

Objective: Blood cell trauma in cardiac surgery is mainly attributed to the use of the heart lung machine and the oxygenator.The optimization of oxygenator design and membrane technology (e.g., chemical properties of surface material), aiming at reducing the blood cell alterations, is an objective of current developments.Clinical evaluation of different oxygenators concerning alterations of blood cells and markers of organ function provide an appropriate level of complexity because various patient characteristics and operative data are involved.Our clinical research database enables us to study the impact of different oxygenator types on blood cells and markers of organ function in the homogenous group of elective coronary artery bypass grafting (CABG) patients.Materials and Methods: Prospective data from 4847 patients who underwent isolated CABG were collected, of which 282 had off-pump coronary artery bypass grafting (OPCAB).The blood cell parameters white blood cell (WBC) counts, platelet (PLT) counts, and red blood cell counts and the marker of organ function lactate dehydrogenase (LDH) and creatinine clearance (CC) were determined pre-and postoperatively.The changes of these parameters were analyzed among the 13-oxygenator type groups and compared with the OPCAB group.To account for individual differences in blood volume, flow rate, and bypass time, we introduced a formula to normalize blood alterations per transfer cycle across the oxygenator.Results: We could observe significant different blood cell alterations between the various oxygenator groups.The increase of WBC was positive correlated with LDH and negative with CC.The oxygenators with the least decrease in PLT had the highest WBC increase.Two manufacturer's modifications in oxygenator surface material are observed to result in a trade off between induced WBC and PLT blood alterations. Conclusion:Cardiotomy suction blood separation had a clear effect on blood activation parameters.The cardiotomy suction blood separation enabled us to separate a main part of blood activation and prevent its negative effects.Further studies should prove whether retransfusion of cardiotomy suction blood without cell saver is indicated at all in routine CABG.
BibTeX
@article{arnrich2004rd,
  author = {Bert Arnrich and A. Albert and Johanna Walter and Frank Schön and M Rothe and Jürgen Ennker and Amir Khosravi and Andreas Liebold and M.S. Joharchi and K. David Wegner and Bernd Westphal and Gustav Steinhoff},
  title = {{33rd International Meeting of the German Society for Cardiovascular Engineering}},
  journal = {Journal of ExtraCorporeal Technology},
  year = {2004},
  doi = {10.1051/ject/2004363286},
}
Figure