128-slice Dual-source Computed Tomography Coronary Angiography in Patients with Atrial Fibrillation: Image Quality and Radiation Dose of Prospectively Electrocardiogram-triggered Sequential Scan Compared with Retrospectively Electrocardiogram-gated Spiral Scan

Lin Lü, Yining Wang, Lingyan Kong, Zhengyu Jin, Guangming Lu, Zhaoqi Zhang, Jian Cao, Shuo Li, Lan Song, Zhiwei Wang, Kang Zhou, Ming WangChinese 2012 · DOI

Abstract

Objective To evaluate the image quality (IQ) and radiation dose of 128-slice dual-source computed tomography (DSCT) coronary angiography using prospectively electrocardiogram (ECG)-triggered sequential scan mode compared with ECG-gated spiral scan mode in a population with atrial fibrillation. Methods Thirty-two patients with suspected coronary artery disease and permanent atrial fibrillation referred for a second-generation 128-slice DSCT coronary angiography were included in the prospective study. Of them, 17 patients (sequential group) were randomly selected to use a prospectively ECG-triggered sequential scan, while the other 15 patients (spiral group) used a retrospectively ECG-gated spiral scan. The IQ was assessed by two readers independently, using a four-point grading scale from excel-lent (grade 1) to non-assessable (grade 4), based on the American Heart Association 15-segment model. IQ of each segment and effective dose of each patient were compared between the two groups. Results The mean heart rate (HR) of the sequential group was 96±27 beats per minute (bpm) with a variation range of 73±25 bpm, while the mean HR of the spiral group was 86±22 bpm with a variationrange of 65±24 bpm. Both of the mean HR (t=1.91, P=0.243) and HR variation range (t=0.950, P=0.350) had no significant difference between the two groups. In per-segment analysis, IQ of the sequential group vs. spiral group was rated as excellent (grade 1) in 190/244 (78%) vs. 177/217 (82%) by reader1 and 197/245 (80%) vs. 174/214 (81%) by reader2, as non-assessable (grade 4) in 4/244 (2%) vs. 2/217 (1%) by reader1 and 6/245 (2%) vs. 4/214 (2%) by reader2. Overall averaged IQ per-patient in the sequential and spiral group showed equally good (1.27±0.19 vs. 1.25±0.22, Z=-0.834, P=0.404). The effective radiation dose of the sequential group reduced significantly compared with the spiral group (4.88±1.77 mSv vs. 10.20±3.64 mSv; t=-5.372, P=0.000). Conclusion Compared with retrospectively ECG-gated spiral scan, prospectively ECG-triggered sequential DSCT coronary angiography provides similarly diagnostically valuable images in patients with atrial fibrillation and significantly reduces radiation dose.
BibTeX
@inproceedings{lu2012slice,
  author = {Lin Lü and Yining Wang and Lingyan Kong and Zhengyu Jin and Guangming Lu and Zhaoqi Zhang and Jian Cao and Shuo Li and Lan Song and Zhiwei Wang and Kang Zhou and Ming Wang},
  title = {{128-slice Dual-source Computed Tomography Coronary Angiography in Patients with Atrial Fibrillation: Image Quality and Radiation Dose of Prospectively Electrocardiogram-triggered Sequential Scan Compared with Retrospectively Electrocardiogram-gated Spiral Scan}},
  booktitle = {Chinese},
  year = {2012},
  doi = {10.1016/s1001-9294(13)60001-2},
}

Loading related content...

Figure