RETRACTED ARTICLE: Implantable neural probes with monolithically integrated CNTFET arrays for multimodal monitoring

Jie Xia, Luxi Zhang, Shengming Wang, Yanlan Yu, Li Ding, Fan Zhang, Shaomin Zhang, Jikui Luo, Yan Yan Shery Huang, Luigi G. Occhipinti, Gang Pan, Zhen Cao, Guoqing Ding, Shurong DongNature Communications 2025 · DOI · OpenAlex

Abstract

The development of implantable neural probes for the simultaneous recording of diverse brain signals remains a critical challenge in neuroscience and clinical diagnostics. This paper introduces an implantable neural probe featuring monolithically integrated carbon nanotube field-effect transistors (CNTFETs), capable of multimodal recording of both electrical and chemical signals under high-field magnetic resonance imaging (MRI). We demonstrate simultaneous measurement of electrophysiological signals with a high signal-to-noise ratio (SNR) of up to 40.34 dB and calcium ion ($Ca^{2+}$) concentrations with a detection limit as low as 0.47 nM. In vivo studies in rat models reveal that changes in $Ca^{2+}$ concentration occur concurrently with epileptiform local field potential (LFP) events, suggesting a robust alternative for accurate epilepsy detection. This work provides a powerful platform for future brain studies and holds significant potential for practical diagnostic applications.

Introduction

Understanding the complex dynamics of the brain requires the simultaneous tracking of electrical activities—such as action potentials and local field potentials (LFPs)—and chemical fluctuations of ions and neurotransmitters. Traditional passive neural probes, typically composed of thin metallic wires, often suffer from significant signal degradation and susceptibility to external electromagnetic interference during transmission to external amplifiers. Furthermore, the presence of bulk metals in these probes poses a significant challenge for Magnetic Resonance Imaging (MRI), leading to severe image artifacts and potential tissue heating due to eddy currents.

To address these challenges, we present the Monolithically Integrated Carbon Nanotube Field-Effect Transistor Array Neural Probe (MICA-NP). This "active" sensing platform leverages the unique properties of carbon nanotubes (CNTs) to provide high-fidelity, multimodal recording of brain signals with exceptional MRI compatibility.

Device Architecture and Fabrication

The MICA-NP is fabricated using a wafer-scale monolithic process, allowing for the integration of 178 probes on a single 4-inch silicon wafer. Each probe features an array of CNTFETs designed for specific sensing modalities. The substrate consists of Si/SiO2, with Y2O3/HfO2 serving as the gate insulator. The source, drain, and gate electrodes are composed of Ti/Pd/Au, while polyimide is used for passivation. This layered structure ensures both electrical performance and mechanical flexibility within the brain environment.

Schematic of the MICA-NP in vivo and its layered monolithic structure.

Multimodal Sensing Mechanism

The probe integrates two distinct types of sensors based on the CNTFET architecture. The LFP active electrodes record electrophysiological signals by directly coupling the brain tissue to the transistor gate. In contrast, the calcium sensors function as bio-field-effect transistors (bioFETs), where the gate is functionalized with a calcium ion-selective membrane (ISM), specifically using the ionophore ETH 5234. This dual-mode design enables the simultaneous acquisition of electrical and chemical data from the same neural environment, with on-site amplification reducing noise susceptibility.

Circuit diagram and sensing principles for LFP and calcium ion detection.

Results and Discussion

Electrical and Chemical Performance

The MICA-NP demonstrates superior performance in both sensing domains. The LFP electrodes achieved a high SNR of 40.34 dB with a wide acquisition bandwidth (0.01 Hz to 32 kHz). The calcium sensors exhibited remarkable sensitivity, with a detection limit of 0.47 nM, significantly outperforming many existing bioFET technologies. In vitro benchmarking confirmed high selectivity for $Ca^{2+}$ over other common ions like $Na^+$ and $K^+$, as well as stability across different pH levels.

MRI Compatibility and Safety

A key advantage of the MICA-NP is its "invisibility" under MRI. By utilizing diamagnetic CNTs and minimizing the thickness of metallic interconnections, the probe's magnetic susceptibility is closely matched to that of brain tissue. Under a 7.0 Tesla magnetic field, the probe produced negligible image artifacts and a peak temperature rise of less than 0.15 °C, well within safety limits and significantly better than traditional platinum electrodes (~0.3 °C).

In Vivo Validation and Longevity

Long-term biocompatibility studies in Sprague-Dawley rats showed minimal chronic inflammation over 12 weeks, evidenced by low astrocyte and microglia activation. The probe remained functional for 3 months in vivo, although the SNR naturally declined from ~40 dB to ~30 dB over time due to the formation of a glial scar. In an acute epilepsy model, the MICA-NP successfully recorded concurrent electrical spikes and calcium concentration changes. Notably, calcium fluctuations were observed to occur concurrently with (or slightly before) electrical events, suggesting that chemical monitoring could provide a more sensitive or earlier detection method for neurological events than electricity alone.

Conclusion

The MICA-NP represents a significant advancement in neuroengineering, providing a stable, active, and MRI-compatible platform for multimodal brain monitoring. While future work will focus on integrating a wider variety of sensors and further miniaturizing the posterior circuitry, this technology offers a powerful tool for both basic neuroscience research and potential clinical diagnostic applications.

Summary updated

Abstract

The implantable neural probe for simultaneous recording of various brain signals is one of the key technologies for neurological science and clinics that is yet to be broken through. Here, we introduce an implantable neural probe with integrated carbon nanotube field-effect transistors which is able to perform multimodal recording of electrical and chemical signals of the brain under magnetic resonance imaging (MRI). We demonstrate here a simultaneous measurement of an electrophysiological signal with high signal-to-noise ratio up to 40.34 dB and calcium concentration with a detection limit down to 0.47 nM. We use our neural probes to detect neural activity in rats and results reveal that changes in Ca²⁺ concentration occur concurrently with the epileptiform local field potential events, providing an alternative method for accurate detection of epilepsy. Our work may provide a powerful means for the future studies of brain and holds great potential for practical diagnostic applications. Multimodal brain recordings are no easy task. Here, the authors introduce a neural implant with integrated carbon nanotube field-effect transistors for multimodal recording of electrical and chemical signals.
BibTeX
@inproceedings{xia2025retracted,
  author = {Jie Xia and Luxi Zhang and Shengming Wang and Yanlan Yu and Li Ding and Fan Zhang and Shaomin Zhang and Jikui Luo and Yan Yan Shery Huang and Luigi G. Occhipinti and Gang Pan and Zhen Cao and Guoqing Ding and Shurong Dong},
  title = {{RETRACTED ARTICLE: Implantable neural probes with monolithically integrated CNTFET arrays for multimodal monitoring}},
  booktitle = {Nature Communications},
  year = {2025},
  doi = {10.1038/s41467-025-67535-5},
}
Figure