Introduction
The study of superheavy element synthesis relies heavily on the precise detection of evaporation residues (EVRs)—heavy nuclei produced when a projectile fuses with a target. At the VASSILISSA electrostatic separator in Dubna, detecting these residues presents significant technical hurdles. EVRs are characterized by low kinetic energies ($0.05–0.2$ MeV/u) and high masses ($A > 200$). These properties lead to a "pulse height defect" in standard silicon detectors, complicating energy measurements. Furthermore, the residues are distributed over a large focal plane area ($25–50$ cm²), and the signal is often obscured by background radiation and scattered beam projectiles. To address these issues, a high-efficiency Time-of-Flight (TOF) system was developed to provide velocity measurements and background rejection.
System Design and Methodology
The TOF system consists of two identical timing detectors placed 40 cm apart. The detection principle relies on secondary electron (SE) emission. As a heavy ion traverses a thin emitter foil (Formvar coated with gold, $20–30 \mu\text{g/cm}^2$), it ejects secondary electrons.
To ensure high timing precision, the system utilizes isochronous transport, where electrons are accelerated and guided by a magnetic field toward Microchannel Plates (MCPs). A key innovation in this design is the use of a 90° image rotation for electron transport. While traditional designs often use a 180° cycloidal path, the 90° approach allows for a more compact magnet configuration suitable for the large active area required at VASSILISSA. By collecting SEs from both sides of the foil and employing four separate MCP multipliers, the system achieves redundancy that ensures near-total detection efficiency for heavy ions.
Experimental Performance
The system's performance was evaluated using both calibration sources and real-world fusion reactions. Initial tests with a $^{226}\text{Ra}$ alpha source demonstrated an intrinsic time resolution of $380 \pm 30$ ps for a single detector and a registration efficiency of approximately 95%.
In heavy ion reaction tests, such as $^{22}\text{Ne} + ^{182}\text{W} \rightarrow ^{204}\text{Po}^*$ and $^{40}\text{Ar} + ^{164}\text{Dy} \rightarrow ^{204}\text{Po}^*$, the system's efficiency for heavy residues exceeded 99.99%. This near-perfect efficiency is attributed to the high number of secondary electrons produced by heavy, slow-moving ions compared to alpha particles. The full two-detector system achieved a timing resolution between 560 and 600 ps, providing a mass resolution of approximately 5%. This resolution is sufficient to distinguish heavy EVRs from lighter background particles like scattered Oxygen or Nickel ions.
Discussion and Conclusion
The TOF system significantly enhances the sensitivity of the VASSILISSA separator by acting as an effective veto. By correlating TOF signals with events in the final silicon detector, researchers can filter out background noise. For instance, if a decay event is recorded in the silicon detector without a coincident TOF "start" signal, it can be confidently identified as the decay of a previously implanted nucleus rather than a new incoming particle.
While the 5% mass resolution is "rough" and cannot yet distinguish between individual neutron evaporation channels (e.g., 3n vs 4n), it provides a vital tool for identifying heavy residues in high-background environments. Future refinements, such as optimizing grid transparency, may further reduce ion scattering and improve performance. This system represents a robust solution for the challenges of large-area, low-energy heavy ion detection in experimental nuclear physics.
Reference: Andreyev, A. N., et al. (1994). Large area high-efficiency time-of-flight system for detection of low energy heavy evaporation residues at the electrostatic separator VASSILISSA. JINR Preprint E15-94-471.