Commissioning of the vacuum system of the KATRIN Main Spectrometer

M. Arenz, M. Babutzka, Matthew Bahr, John Barrett, Stephan Bauer, M. Beck, A. Beglarian, J. Behrens, Thaisa Storchi‐Bergmann, U. Besserer, J. Blümer, Laura Bodine, K. Bokeloh, J. Bonn, B. Bornschein, L. Bornschein, S. Büsch, T. H. Burritt, S. Chilingaryan, T.j. Corona, L. de Viveiros, P. J. Doe, O. Dragoun, G. Drexlin, S. Dyba, S. Ebenhöch, K. Eitel, E. Ellinger, S. Enomoto, M. Erhard, D. Eversheim, M. Fedkevych, A. Felden, S. Fischer, J. A. Formaggio, F. M. Fränkle, D. Furse, M. C. Ghilea, W. Gil, F. Glück, A. González Ureña, S. Görhardt, Stefan Groh, S. Grohmann, Robin Größle, R. Gumbsheimer, M. Hackenjos, V. Hannen, F. Harms, N. Haußmann, F. Heizmann, K. Helbing, Werner Herz, S. Hickford, D. Hilk, B. Hillen, T. Höhn, B. Holzäpfel, M. Hötzel, Howe, A. Huber, A. Jansen, N. Kernert, L. Kippenbrock, M. Kleesiek, Manuel Klein, A. Kopmann, A. Kosmider, A. Kovalík, B. Krasch, M. Kraus, H. Krause, M. Krause, L. Kuckert, B. Kuffner, L. La Cascio, O. Lebeda, B. Leiber, J. Letnev, V.m. Lobashev, Alexey Lokhov, E. Malcherek, M. Mark, E. L. Martín, S. Mertens, S. Mirz, B. Monreal, K. Müller, M. Neuberger, H. Neumann, S. Niemes, M. Noë, N. S. Oblath, A. Off, H.-w. Ortjohann, A. Osipowicz, E. Otten, D. S. Parno, P. Plischke, A.w.p. Poon, R. M. D. Carney, M. Descher, F. Edzards, F. Friedel, A. S. Gavin, C. KarlRepository 2016 · DOI · arXiv · OpenAlex

Abstract

The KATRIN experiment will probe the neutrino mass by measuring the beta-electron energy spectrum near the endpoint of tritium beta-decay. An integral energy analysis will be performed by an electro-static spectrometer (Main Spectrometer), an ultra-high vacuum vessel with a length of 23.2 m, a volume of 1240 m^3, and a complex inner electrode system with about 120,000 individual parts. The strong magnetic field that guides the beta-electrons is provided by super-conducting solenoids at both ends of the spectrometer. Its influence on turbo-molecular pumps and vacuum gauges had to be considered. A system consisting of 6 turbo-molecular pumps and 3 km of non-evaporable getter strips has been deployed and was tested during the commissioning of the spectrometer. In this paper the configuration, the commissioning with bake-out at 300°C, and the performance of this system are presented in detail. The vacuum system has to maintain a pressure in the $10^{-11}$ mbar range. It is demonstrated that the performance of the system is already close to these stringent functional requirements for the KATRIN experiment, which will start at the end of 2016.

Introduction

The Karlsruhe Tritium Neutrino (KATRIN) experiment aims to determine the absolute mass scale of neutrinos with a sensitivity of 200 meV. This is achieved by precisely measuring the endpoint of the tritium beta-decay spectrum. To reach this sensitivity, the experiment requires an unprecedented level of background suppression and energy resolution. Central to this is the Main Spectrometer (MS), a massive ultra-high vacuum (UHV) vessel that acts as a high-pass filter for electrons.

The KATRIN experimental setup, showing the beamline from the source (WGTS) to the detector (FPD).

The MS is a 23.2-meter-long stainless steel vessel with a volume of 1,240 m³. To minimize background signals from collisions between signal electrons and residual gas molecules, the spectrometer must maintain a base pressure in the $10^{-11}$ mbar range—roughly 14 orders of magnitude lower than atmospheric pressure.

The Vacuum System and Spectrometer Design

The spectrometer operates on the principle of Magnetic Adiabatic Collimation combined with an Electrostatic (MAC-E) filter. Electrons from the tritium source are guided by magnetic fields through the spectrometer.

Schematic of the Main Spectrometer showing electron tracks, the inner electrode system, and the magnetic field configuration.

The vacuum system is composed of several critical components:

  • Pumping Hardware: Six turbo-molecular pumps (TMPs) handle the initial pump-down. For the UHV regime, three custom pump ports house 3 km of SAES St707 non-evaporable getter (NEG) strips, which provide a massive pumping speed for hydrogen.
  • Inner Electrodes: A complex system of 120,000 individual parts forms the dual-layer electrode system used to create the retarding potential.
  • Bake-out System: To achieve UHV, the entire vessel was baked at 300°C for several weeks using a 440 kW thermal oil system to desorb water and other contaminants from the inner surfaces.

Commissioning and Key Results

The commissioning phase involved a rigorous bake-out cycle and testing of the pumping systems. Despite challenges, including mechanical deformations of electrode components due to thermal stress at 300°C, the system demonstrated exceptional performance.

Vacuum Performance

The system successfully reached an ultimate pressure in the $10^{-11}$ mbar range, dominated by hydrogen. The measured hydrogen outgassing rate was $1.4 \times 10^{-12}$ mbar·l/s·cm², which is remarkably low for a vessel of this scale and complexity. The effective pumping speed for hydrogen was maintained at 300 m³/s, even with the NEG pumps operating at partial activation to manage thermal loads.

Technical Insights

A key finding during commissioning was the resilience of the NEG pumps. When a valve failure necessitated venting the vessel, the team used ultra-pure Grade 9.0 argon. This prevented the "poisoning" of the NEG strips, allowing the system to recover without a full reactivation cycle. Additionally, liquid nitrogen (LN2) cooled baffles were confirmed to be effective at trapping Radon atoms, which are a significant source of background noise.

Conclusion and Outlook

The commissioning of the KATRIN Main Spectrometer vacuum system has proven that achieving and maintaining $10^{-11}$ mbar in a 1,240 m³ vessel is feasible. While thermal stress caused some mechanical damage to the internal wire electrodes, mitigation strategies have allowed the experiment to proceed. Following these tests, the spectrometer was integrated into the full beamline, paving the way for high-precision neutrino mass measurements.

Summary updated

Abstract

The KATRIN experiment will probe the neutrino mass by measuring the beta-electron energy spectrum near the endpoint of tritium beta-decay. An integral energy analysis will be performed by an electro-static spectrometer (Main Spectrometer), an ultra-high vacuum vessel with a length of 23.2 m, a volume of 1240 m^3, and a complex inner electrode system with about 120000 individual parts. The strong magnetic field that guides the beta-electrons is provided by super-conducting solenoids at both ends of the spectrometer. Its influence on turbo-molecular pumps and vacuum gauges had to be considered. A system consisting of 6 turbo-molecular pumps and 3 km of non-evaporable getter strips has been deployed and was tested during the commissioning of the spectrometer. In this paper the configuration, the commissioning with bake-out at 300°C, and the performance of this system are presented in detail. The vacuum system has to maintain a pressure in the 10^{-11} mbar range. It is demonstrated that the performance of the system is already close to these stringent functional requirements for the KATRIN experiment, which will start at the end of 2016.
BibTeX
@inproceedings{arenz2016commissioning,
  author = {M. Arenz and M. Babutzka and Matthew Bahr and John Barrett and Stephan Bauer and M. Beck and A. Beglarian and J. Behrens and Thaisa Storchi‐Bergmann and U. Besserer and J. Blümer and Laura Bodine and K. Bokeloh and J. Bonn and B. Bornschein and L. Bornschein and S. Büsch and T. H. Burritt and S. Chilingaryan and T.J. Corona and L. de Viveiros and P. J. Doe and O. Dragoun and G. Drexlin and S. Dyba and S. Ebenhöch and K. Eitel and E. Ellinger and S. Enomoto and M. Erhard and D. Eversheim and M. Fedkevych and A. Felden and S. Fischer and J. A. Formaggio and F. M. Fränkle and D. Furse and M. C. Ghilea and W. Gil and F. Glück and A. González Ureña and S. Görhardt and Stefan Groh and S. Grohmann and Robin Größle and R. Gumbsheimer and M. Hackenjos and V. Hannen and F. Harms and N. Haußmann and F. Heizmann and K. Helbing and Werner Herz and S. Hickford and D. Hilk and B. Hillen and T. Höhn and B. Holzäpfel and M. Hötzel and Howe and A. Huber and A. Jansen and N. Kernert and L. Kippenbrock and M. Kleesiek and Manuel Klein and A. Kopmann and A. Kosmider and A. Kovalík and B. Krasch and M. Kraus and H. Krause and M. Krause and L. Kuckert and B. Kuffner and L. La Cascio and O. Lebeda and B. Leiber and J. Letnev and V.M. Lobashev and Alexey Lokhov and E. Malcherek and M. Mark and E. L. Martín and S. Mertens and S. Mirz and B. Monreal and K. Müller and M. Neuberger and H. Neumann and S. Niemes and M. Noë and N. S. Oblath and A. Off and H.-W. Ortjohann and A. Osipowicz and E. Otten and D. S. Parno and P. Plischke and A.W.P. Poon and R. M. D. Carney and M. Descher and F. Edzards and F. Friedel and A. S. Gavin and C. Karl},