The Motion-Driven Distortion Floor in Radar Vital-Sign Sensing

Radar can measure a heartbeat from across a room by tracking sub-millimeter movements of the chest wall. The approach fails once the person walks, turns, or shifts in a chair. The field's working explanation is that body motion is a large interfering signal that a clever enough network can subtract away. This note proposes a different explanation for part of that failure. A moving torso or limb returns a radar echo that is orders of magnitude stronger than the cardiac echo. Every real receiver adds distortion in proportion to the power it handles, through amplifier nonlinearity, oscillator phase noise, imbalance between its in-phase and quadrature channels, and analog-to-digital converter (ADC) errors. The strong motion echo therefore carries its own distortion, and that distortion falls on top of the heartbeat. No transmit power, antenna count, or downstream network trained on the digitized signal can fully remove noise that the hardware itself generated. Adding power scales both echoes and their distortion together.

We propose to measure this floor, predict it from a small number of per-device hardware constants, and test two ways of acting on it. The first is a whitening step placed before the neural network. It estimates how the motion-induced distortion is spread across range bins and antennas, then suppresses it. The second is a reliability estimate that tells the system when a window lies below the floor and should be left unreported. The central claim is falsifiable. Cardiac displacement error under motion should flatten into a floor set by the ratio of cardiac to motion echo power and a device constant $\kappa$, and that floor should not move with transmit power. If the claim holds, much of what the literature calls "motion artifact" is really a hardware-limited signal-to-distortion problem, and the benchmarks should report it that way.

Core Claim

Communications theory has long modeled residual transceiver impairments as additive Gaussian noise whose power scales with signal power. The signal-to-noise-and-distortion ratio then becomes $P/(\sigma^2+\kappa P)$ and saturates at $1/\kappa$. MIMO Transmission with Residual Transmit-RF Impairments validated the Gaussian model on a multi-antenna testbed. A New Look at Dual-Hop Relaying: Performance Limits with Hardware Impairments derived the resulting outage and capacity ceilings. Full-Duplex Bidirectional MIMO: Achievable Rates Under Limited Dynamic Range showed that a strong self-interferer, together with limited receiver dynamic range, produces a floor that neither cancellation nor more training pilots can beat. Beyond roughly 50 pilots, training saturates.

The radar vital-sign problem has the same structure, with the macro-motion echo playing the role of the self-interferer. Let $P_c$ be the received power of the cardiac micro-motion component and $P_m$ the power of the macro-motion echo. Let $\sigma^2$ be thermal noise and $\kappa$ an effective device constant that aggregates amplifier, phase-noise, imbalance, and ADC distortion. The effective ratio for the cardiac component is then

$$ \mathrm{SNDR}_c = \frac{P_c}{\sigma^2 + \kappa\,(P_m + P_c)} \;\xrightarrow{\;P_{\text{tx}}\to\infty\;}\; \frac{\rho}{\kappa(1+\rho)}, \qquad \rho = P_c/P_m . $$

Transmit power scales $P_c$ and $P_m$ together, so the limit depends only on the geometry ratio $\rho$ and the hardware constant $\kappa$. The standard phase-estimation bound gives displacement variance $\approx \lambda^2/(32\pi^2 N_{\text{eff}}\,\mathrm{SNDR}_c)$. Here $\lambda$ is the carrier wavelength, and the formula follows because displacement is $\lambda/(4\pi)$ times phase and phase variance for a tone in noise is about $1/(2N_{\text{eff}}\,\mathrm{SNDR})$. The quantity $N_{\text{eff}}$ is the number of effectively independent samples, which is reduced when the distortion is colored. The resulting prediction is a displacement-error floor that rises as the motion echo grows and ignores the transmit power knob.

Performance Limits of Hardware-Constrained THz Inter-Satellite MIMO-ISAC Systems adds a second prediction. Amplifier distortion beamforms coherently with the probe signal, which is why the critical SNR there was scale-invariant in array size (slope ≈ 0). Adding antennas should therefore not lower the floor either, although it may improve ranging. That paper motivates the array-scaling test. Its setting, inter-satellite tracking, is otherwise far from human physiology.

What Is and Is Not New

Radar vital-sign systems currently treat body motion as a signal-level interferer to be learned away. RF-vital, in Remote Respiration Monitoring of Moving Person Using Radio Signals, trains adversarial decoders so that walking-trajectory features do not leak into the breathing representation. MoRe-Fi: Motion-robust and Fine-grained Respiration Monitoring via Deep-Learning UWB Radar processes raw complex in-phase/quadrature data with a variational encoder-decoder. Its motivation is that projecting onto amplitude or phase alone distorts the signal. V2iFi: in-Vehicle Vital Sign Monitoring via Compact RF Sensing abandons phase for amplitude because phase became unstable under vehicle vibration. That is consistent with, though not proof of, a phase-noise-driven floor. Contactless 12-lead electrocardiogram via deep computational radar uses sub-\$30 60 GHz hardware and reports that gross body movement breaks reconstruction.

All four papers motivate this work, because each documents motion failures. None models the receiver as a nonideal device whose distortion scales with the motion echo. Conversely, the hardware-impairment papers supply the method: the proportional-distortion model, and the whitening-filter idea from Studer and colleagues, which recovered about 6 dB at 1% frame error rate. They never consider a physiological target or a strong moving reflector. This proposal does three things neither line of work does. It measures a per-device motion-driven floor. It tests whether learned disentanglement is bounded by it. It checks whether hardware-aware whitening moves learned models closer to it.

Research Program

We first build a controlled testbed around two commercial radars, one 60 GHz FMCW and one impulse-radio ultra-wideband, chosen to span hardware quality. Each radar's receiver gain, transmit power, and ADC bit depth are software-configurable; where bit depth is not configurable, we emulate it by truncating bits. The test target is a mechanical phantom: a voice-coil actuator produces calibrated sub-millimeter "cardiac" vibration while mounted on a linear stage that produces centimeter-scale "gait" motion at 1 to 2 Hz. This gives exact ground truth and a directly controllable $\rho$. We estimate $\kappa$ for each device independently of the vital-sign task. The procedure drives the receiver with a strong corner-reflector echo at known motion and measures residual error vector magnitude in the empty range bins nearby, mirroring how communications work characterizes impairments.

Second, we derive the floor and its structure. Because distortion from a single strong reflector enters the receiver through known paths, its covariance across range bins and antennas can be estimated from the motion echo itself. That covariance determines $N_{\text{eff}}$, and it is what a whitening filter would exploit. Oscillator phase noise is partly cancelled for nearby targets, since the transmitted and received signals share the same oscillator. We model the residual as growing with the range offset between torso and chest, so $\kappa$ is split into a range-independent part and a phase-noise part.

Third, we build the intervention. For each window, a covariance estimate is taken from range bins dominated by the motion echo, and a whitening transform in the style of the QR-based whitening of Studer and colleagues is applied to the complex in-phase/quadrature cube. The result feeds an unchanged MoRe-Fi-style encoder. From the same estimate, the model computes predicted $\mathrm{SNDR}_c$. A window is abstained on when its predicted bound exceeds a clinical tolerance, for example 5 bpm heart-rate error.

Fourth, we move to humans. Twelve to twenty subjects perform sitting, typing, turning, and treadmill walking at several speeds, with reference ECG and a respiration belt, at three transmit powers and three receiver gains. The protocol mirrors the activity sets of MoRe-Fi and RF-vital, so our numbers can be read against theirs.

Baselines

The naive reference is a standard pipeline with static-clutter subtraction, a range-bin selection step, and a heart-rate band-pass filter at 0.8 to 3 Hz, run at default gain. It sets the level that any claimed improvement must clear.

The strongest existing learned competitor is a reimplementation of MoRe-Fi's complex in-phase/quadrature variational encoder-decoder, extended with a cardiac head. It is the only published method that consumes raw complex data, and it is therefore the most plausible model to absorb hardware distortion implicitly. MoRe-Fi reports an average cosine similarity of 0.9162 against belt waveforms for respiration on its own dataset (MoRe-Fi). RF-vital reports a walking respiration mean absolute error of 3.67 BPM (Remote Respiration Monitoring of Moving Person Using Radio Signals). For heart rate under vibration, V2iFi reports a median error of 0.6 bpm in driving (V2iFi). These figures come from different hardware and populations, so they serve as sanity anchors only. All comparisons will be on our data, with our reimplementation retrained per device, identical splits by subject, the same compute budget, and the same hyperparameter search for every variant. We found no reported result on Lacuna for learned radar heart-rate estimation during treadmill walking, so we cannot quote a walking heart-rate number.

We also include an adversarial disentanglement baseline in the style of RF-vital, which uses a motion decoder penalty, and a classical multi-bin decomposition in the style of V2iFi's multi-sequence variational mode decomposition on amplitude.

Three ablations isolate the key ingredient. The first is the same encoder trained with simulated proportional-distortion augmentation but no whitening. The second is whitening using a mis-specified, isotropic covariance. The third is abstention driven by a learned quality classifier in the spirit of Tiny-PPG rather than by the physical bound. Beating the strongest baseline means two things. At matched coverage, the whitened model must lower heart-rate error in walking windows by a margin outside subject-level bootstrap intervals. At matched error, physics-based abstention must retain more windows than the learned quality detector.

Why not just…?

Why not just turn the receiver gain down, or use automatic gain control, so nothing saturates? This is a real threat, and we treat it as an experiment. Backing off gain reduces the distortion term but raises the relative thermal term, so our equation predicts an optimal gain that depends on $\rho$ rather than a free fix. Proportional distortion from phase noise and in-phase/quadrature imbalance also persists without any clipping, which is the core of the model in Studer and colleagues. We sweep gain on every device. If a tuned gain schedule alone closes most of the gap to the floor, we will say so, and the contribution narrows to predicting that schedule from $\kappa$.

Why not just boost transmit power or add antennas? The equation says power cancels out, and the THz ISAC analysis found the critical SNR invariant to array size because distortion beamforms with the signal (Performance Limits of Hardware-Constrained THz Inter-Satellite MIMO-ISAC Systems). Our power sweep and the 3×4 versus single-channel comparison test this directly. If error keeps falling with power under motion, the floor claim is wrong.

Why not just train the network on data corrupted with simulated impairments and let it learn? This is the strongest alternative. It fits the "noise as an asset" position of Pulse-PPG and A robust PPG foundation model using multimodal physiological supervision. Amplifier and ADC nonlinearities are deterministic functions of the input, so a network could partially invert them. Phase noise is random, though, and Studer and colleagues found that sophisticated detectors degrade more than linear ones when colored hardware noise goes unmodeled. The augmentation-only ablation settles the question. If it matches whitening, we report that the floor is real but learnable around by augmentation, and the whitening step is unnecessary.

Why not just detect motion windows and discard them? Discarding costs coverage exactly when monitoring matters, such as during exercise or in patients who move. A learned detector, as in Tiny-PPG with its 87.4% DICE on PPG-DaLiA, flags artifacts but cannot say whether a window is recoverable. The matched-coverage comparison tests whether the physical bound makes that distinction better.

Decisive Checks

The first check comes from the phantom. At fixed $\rho$, cardiac displacement error across a tenfold transmit-power sweep should flatten to within bootstrap error of the $\kappa$-predicted floor. The floor should shift by the predicted amount when $\rho$ is changed through stage amplitude. The cheaper device should show a proportionally higher floor. Failure on any of these falsifies the core model.

The second check is whether all learned baselines' walking-window errors on humans sit on or above the predicted floor when binned by estimated $\mathrm{SNDR}_c$. Systematic sub-floor performance would mean the network exploits structure the Gaussian model ignores, which would itself be a useful finding.

The third check is whether whitening moves the MoRe-Fi-style model toward the floor beyond what augmentation alone achieves.

The fourth check is whether bound-based abstention dominates learned quality gating on the coverage-versus-error curve.

As a secondary, low-cost probe of the same mechanism in wearables, we will re-quantize PPG-DaLiA wrist signals through the 4-bit SAR-ADC mismatch model of The Impact of SAR-ADC Mismatch on Quantized Massive MU-MIMO Systems, which reported roughly a 1% uncoded bit-error floor at half-LSB mismatch in the MIMO setting. We then measure whether motion-induced swings produce heart-rate error floors for a fixed PPG model. This is simulation-only and separate from the radar claim.