Automatic radiology report generators for 3D CT are trained almost entirely on scans cut into thin slices, a millimetre or so apart. Many hospitals, especially community centres, acquire routine CT with slices 3 to 5 mm thick. A lesion smaller than the slice thickness gets averaged into its surroundings and can vanish from the image. A report generator deployed on such scans then faces a choice it was never trained to make. It can write "no nodule," which may be false and is medicolegally dangerous, or it can guess from correlated cues, which is a hallucination. Human radiologists handle this with a third option: they write that a finding "cannot be excluded at this slice thickness." We propose to teach report generators that third option, and to measure how badly current systems fail without it.
The core idea is simple. We take thin-slice chest CTs, compute from each one physically faithful thick-slice versions by averaging adjacent slices, and train the report generator on these paired views. The training target for a thick view is not the original report. It is a relabeled report in which findings that are no longer resolvable at that thickness become explicit indeterminate statements rather than positive or negative assertions. Because the pairs come from the same patient, we can also score exactly which findings a model drops, invents, or wrongly negates as slices thicken. Jeong et al. named this missing ground truth as the main weakness of their own slice-thickness study.
Core Claim
A 3D report generator trained with resolution-aware targets will keep its accuracy on findings that remain visible in thick-slice CT, sharply reduce false negations of findings that became invisible, and replace them with hedges that are calibrated: a hedged finding should be positive on the thin scan much more often than the base rate. We claim this cannot be achieved by resampling, super-resolution preprocessing, or thickness augmentation with the original reports. It may or may not be matched by a post-hoc hedging rule, and we test that directly.
Why This Gap Is Real
The file supports each piece of the motivation. Deep Learning-Based Slice Thickness Reduction for Computer-Aided Detection of Lung Nodules in Thick-Slice CT shows that 5 mm scans hide small nodules through partial-volume averaging and change how nodules are characterized. Its fix is a separate super-resolution network placed in front of a detector, validated without same-patient thin scans. AMOS shows that even multi-centre training leaves drops of over 20% mDice on unseen scanners. Both papers motivate the work, but neither involves language.
On the report side, CT2Rep builds a 3D encoder, with in-plane transformers followed by a causal transformer over the slice stack, and a memory-augmented decoder. It reaches a clinical-efficacy F1 of 0.534 on CT-RATE. CT-CLIP / CT-CHAT pairs a similar encoder with Llama 3.1 and reports gains on external cohorts. BrainGPT generates head CT reports from a 2D CLIP encoder and introduces the FORTE metric. None of these three models the physical spacing between slices. None was tested on thicker acquisitions than it was trained on. All of them produce deterministic reports with no way to say "not assessable."
CSAM shows that uncertainty caused by partial-volume effects in anisotropic scans is learnable: its predicted uncertainty correlates at r = 0.791 with segmentation error. Improved post-hoc probability calibration for out-of-domain MRI segmentation shows that models become overconfident out of domain. Both results are about pixels. Our proposal moves that uncertainty into the report text, where clinicians actually read it.
CARVE is the closest existing work. It exploits redundancy between adjacent anisotropic slices to discard 80% of visual tokens while keeping 98.1% of benchmark performance. It uses heuristic windowing on fixed inputs and tests neither slice-thickness shift nor omission of rare focal findings. We do not propose a better compressor. Instead, CARVE's own premise, that focal lesions live on one or two slices, is exactly what thickening destroys. We use CARVE as a stress case in our evaluation.
Research Program
The first component is a paired-thickness benchmark built from CT-RATE. Using the DICOM spacing metadata, we select volumes reconstructed with thin slices. For each one, we synthesize thick-slice views at several nominal thicknesses (around 2.5, 3.75 and 5 mm) by averaging groups of adjacent slices. We then add mild in-plane noise matched to typical low-dose reconstructions. Slab averaging approximates how scanners reconstruct thicker slices from the same raw data far more closely than learned super-resolution approximates the reverse direction. This lets us hold patient, anatomy and report fixed while varying only through-plane resolution. We also keep a small set of scans where the vendor itself produced both thin and thick reconstructions. This checks that our simulated thick views resemble real ones.
The second component is resolvability labeling. For each abnormality label and each thick view, we decide whether the finding is still visible. We use two signals and require them to agree. The first signal is size: many reports state lesion dimensions in millimetres, and we parse these and compare them to the slice thickness. The operating range around 6 mm is informed by Jeong et al.'s small-nodule stratum. The second signal is a teacher: a frozen CT-CLIP abnormality classifier is run on the thin and thick views of the same scan. A finding that is confidently positive on the thin view but whose probability collapses on the thick view is marked unresolvable. Findings that remain detectable keep their original sentence. Findings that became unresolvable have their sentence rewritten into a templated hedge naming the finding and the slice-thickness limitation. Negative findings in regions where small lesions are plausible, such as "no pulmonary nodule," are softened at large thicknesses to "no nodule larger than X mm." A radiologist audits a random sample of relabeled reports.
The third component is the model. We start from the CT2Rep architecture. We replace the index-based positions of its slice-stack transformer with a continuous encoding of each slice's physical height in millimetres, so the model knows how far apart two slices really are. We then fine-tune on the mixture of thin views with original reports and thick views with relabeled reports. The spacing encoding is a supporting ingredient; the relabeled targets are the main one. The ablations are designed to tell the two apart.
Baselines
The naive reference is CT2Rep trained as published and applied to thick views without any change. On thin CT-RATE it reports a clinical-efficacy F1 of 0.534 and a clinical-efficacy precision of 0.749 CT2Rep. We found no reported result on Lacuna for any CT-RATE report generator on thick-slice inputs. Measuring that degradation curve is itself a first deliverable.
The strongest existing system is CT-CHAT CT-CHAT. Its blinded radiologist review favoured it over CT2Rep and RadFM, and it holds up best of the three on external cohorts. We run the released weights on the same thin and thick test views, with the same generation settings. If we fine-tune it, it gets the same number of steps and the same learning-rate sweep as our model.
The resampling baseline interpolates each thick view back to the thin grid before running the thin-trained model. The super-resolution baseline trains a slice-thickness-reduction network in the style of Jeong et al. on our own thin/thick pairs, giving it privileged same-patient supervision it lacked. It then runs the unchanged generator on the output. The CARVE stress case applies CARVE's token reduction at a 20% keep ratio to thick and thin inputs, to test whether compression and thickening compound omissions.
Two ablations isolate our ingredient. Plain thickness augmentation trains on thick views with the original, unrelabeled reports. Augmentation with the spacing encoding but without relabeling adds only the physical-position change. Every trained variant uses the same backbone, data, epochs and hyperparameter budget.
Beating the baselines means three things together. First, clinical-efficacy F1 on resolvable findings in thick views must be no worse than the best baseline. Second, the false-negation rate on unresolvable findings must fall significantly. We define that rate as the fraction of findings positive on the thin view that the report on the thick view explicitly denies. Third, hedge precision must exceed the base rate of the hedged findings, and the fraction of normal scans receiving hedges must stay low.
Why not just…?
Why not just resample to an isotropic grid? Interpolation adds voxels but no information. E2Net reports that 3D resampling blurs boundaries and inflates memory, and Jeong et al. found simple interpolation insufficient to recover nodule appearance. More fundamentally, a resampled scan still hides the sub-slice lesion, and the downstream model still has no reason to hedge. Our resampling baseline measures this directly through the false-negation rate.
Why not just super-resolve thick slices first? This is a serious competitor, and our version is stronger than any published one because it is trained on same-patient pairs. Its failure mode is structural: a learned upsampler must invent the content of the missing slab. It will therefore either erase small lesions or produce plausible ones, and the report inherits that confident guess. Jeong et al. could not rule out hallucinated structure. We can, by comparing super-resolved reports against the true thin-view findings. If super-resolution matches our false-negation rate without raising hallucinations, we will report that as a positive result for the simpler pipeline.
Why not just augment with thick slices and add a spacing token? This is the ablation that matters most. Training a thick view against a report describing invisible findings teaches the model to assert what it cannot see, which amounts to rewarding hallucination from correlated context. If this ablation matches relabeling on both false negation and hallucination, our central claim is wrong, and the experiment is set up to show exactly that.
Why not just add post-hoc hedging from classifier confidence and slice metadata? This is the most dangerous alternative. One could calibrate a per-finding abnormality head with temperature scaling, in the spirit of Ouyang et al., then insert templated hedges whenever confidence is intermediate and thickness exceeds a threshold. We implement this rule on top of the augmented model and tune its thresholds on validation data. Our trained model has potential advantages: it can learn which anatomical contexts make a finding unresolvable, and it keeps hedges consistent with the rest of the report. If the rule matches it, the contribution reduces to the benchmark and the rule, and we will say so.
Proposed Experiments
The first experiment characterizes the problem. We plot clinical-efficacy F1, the false-negation rate and the hallucination rate against slice thickness for CT2Rep, CT-CHAT and CARVE-compressed variants on the paired CT-RATE test set. We stratify by lesion size wherever reports state it. We expect false negation to rise steeply once thickness passes the parsed lesion size. If it does not, the thickness shift matters less for report generation than the detection literature suggests, and the project narrows to a negative finding.
The second experiment is the main comparison: all baselines and ablations on the paired test set. Its decisive check is the gap between relabeled training and plain augmentation on false negation and hedge precision.
The third experiment tests transfer to real rather than simulated shift. We evaluate on the vendor-reconstructed thin/thick pairs and on RAD-ChestCT, where CT-CLIP's external evaluation provides the reference cohort. For RAD-ChestCT we report label-level clinical-efficacy metrics and hedge rates grouped by the native slice thickness in its metadata.
The fourth experiment is a blinded reader study with a small panel of radiologists. They rate paired reports on thick scans for clinical acceptability and for whether the report would mislead a referring physician. This checks that hedges read as appropriate caution rather than noise.
The whole project is falsifiable on one number. If relabeled targets do not lower the rate at which thick-slice reports deny findings that the thin scan shows, at equal accuracy on visible findings, the idea fails.
References
- Deep Learning-Based Slice Thickness Reduction for Computer-Aided Detection of Lung Nodules in Thick-Slice CT
- AMOS: A Large-Scale Abdominal Multi-Organ Benchmark for Versatile Medical Image Segmentation
- CT2Rep: Automated Radiology Report Generation for 3D Medical Imaging
- Developing Generalist Foundation Models from a Multimodal Dataset for 3D Computed Tomography
- Towards a holistic framework for multimodal LLM in 3D brain CT radiology report generation
- CSAM: A 2.5D Cross-Slice Attention Module for Anisotropic Volumetric Medical Image Segmentation
- Improved Post-hoc Probability Calibration for Out-of-Domain MRI Segmentation
- CARVE: Cross-Slice Anisotropic Reallocation of Visual Evidence for Efficient 3D Medical Volume Understanding
- E2Net: An Edge Enhanced Network for Accurate Liver and Tumor Segmentation on CT Scans