During head-and-neck radiotherapy, the salivary glands receive collateral radiation that often leaves patients with permanent dry mouth (xerostomia). A prospective study found that the glands announce their injury early. By the fifth of roughly thirty-five daily treatments, a diffusion MRI measure of how freely water moves in tissue (the apparent diffusion coefficient, ADC) rises 16–18% in the parotid glands, while gland volume has barely changed. That early rise predicts dry mouth at six months with an area under the curve (AUC) of about 0.77 (Early prediction of parotid glands secretory function based on ADC variations during radiotherapy for nasopharyngeal carcinoma: a phase II prospective study). So far nobody uses this signal to change the treatment. We propose to use it as a measurement of how sensitive each patient's own glands are to dose. With that estimate, a simple budget controller spreads the remaining dose between the tumor, the two parotids, and neighboring organs over the rest of the course.
The control idea comes from wireless spectrum sharing. An LTE base station sharing a channel with Wi-Fi cannot decode Wi-Fi packets. It learns how its unseen neighbor responds only from noisy feedback, and it adjusts how much airtime it takes so that the neighbor is not starved (Wireless Optimisation via Convex Bandits: Unlicensed LTE/WiFi Coexistence). The gland is a similar neighbor. We cannot observe its true tolerance, but each imaging session gives a noisy reading of how it is reacting. Our main claim is narrow and testable. If fraction-5 ADC carries information about gland injury beyond what delivered dose already explains, then a controller that adjusts the price of parotid dose at fraction 5 and again at fraction 15 should lower predicted xerostomia at equal tumor coverage. It should do so relative to static planning, delivered-dose-triggered replanning, and a simple ADC threshold rule. If ADC carries no such extra information, the method has no reason to exist, and we test that first.
What is already known, and what is not
Three studies define the clinical problem. The ADC study above is purely prognostic: it reports cut-offs (ΔADC₅ above 194.1 units ipsilateral, above 179.6 contralateral) but no intervention. A cone-beam CT dose-reconstruction study tracked weekly anatomy with deformable registration. It showed that delivered parotid dose drifts well above plan as glands shrink and migrate, with V40 up 44.4% and mean dose up 9.5%. Delivered dose separated patients who developed xerostomia from those who did not (Parotid gland radiation dose‐xerostomia relationships based on actual delivered dose for nasopharyngeal carcinoma). A static duct-sparing planning study cut excretory-duct dose without losing target coverage, but it slightly raised dose to the brainstem, mandible, and oral cavity. Its authors note that gland migration can move spared regions into high-dose fields (Parotid Gland Stem Cell Preservation during Intensity-Modulated Radiotherapy for Nasopharyngeal Carcinoma: Dosimetric Analysis and Feasibility). Together these motivate in-treatment adaptation, and they warn that sparing one organ pushes dose onto others.
One fact cuts against the naive version of this proposal. In the same ADC cohort, ADC change correlates with delivered dose at r = 0.60–0.68. Much of the "early biomarker" may therefore just re-measure dose, which cone-beam CT already provides. The proposal's value rests on the remaining, patient-specific part of the ADC change.
Automated planning methods supply the optimization machinery but not the feedback loop. One is a deep actor-critic planner that tunes optimization weights offline for cervical cancer (Automatic radiotherapy treatment planning with deep functional reinforcement learning). Another is a multimodal-model planner that picks gantry angles through a simulator loop (Transforming Multimodal Models into Action Models for Radiotherapy). Both plan once, before treatment, on anatomy alone.
On the wireless side, the constrained reinforcement learner QaSAL adds Lagrange multipliers (running "prices" on constraint violations) to the agent's state (QaSAL: QoS-aware State-Augmented Learnable Algorithms for Coexistence of 5G NR-U/Wi-Fi). Two lessons from that literature limit our design rather than supplying it. First, QaSAL guarantees its constraints only on average over time, whereas a patient is a single episode. Second, the convex-bandit method oscillates badly when its exploration step is too large. Both points argue against training a learned policy over many simulated patients. Instead we use a model-based multiplier update with small, conservative step sizes and a per-patient chance constraint.
Core Mechanism
At fraction 5, we register the ADC map to the dose actually accumulated so far, using the deformable-registration method of the delivered-dose study. For each gland we fit a voxel-wise slope $s$ relating ADC change to accumulated dose. A hierarchical Bayesian prior pools these slopes across patients, so each patient gets a posterior over their own gland sensitivity rather than a point estimate. A dose-response model maps $s$ and the projected total dose to a predicted probability of xerostomia, $\hat p_k$, at imaging point $k$.
The planner then re-optimizes the remaining fractions. It minimizes target underdose plus $\lambda_k$ times parotid dose, while keeping hard limits on the brainstem, mandible, and oral cavity. Here $\lambda_k$ is the current price of parotid dose, and it is updated at each imaging point by
$$\lambda_{k+1} = \max\left(0,\; \lambda_k + \eta\,(\hat p_k^{(q)} - B)\right),$$
where $\hat p_k^{(q)}$ is the $q$-th upper quantile of predicted xerostomia probability under the sensitivity posterior, $B$ is the clinician-set toxicity budget, and $\eta$ is a small step size. Using an upper quantile instead of the mean is how we answer QaSAL's average-only guarantee: the budget has to hold for this patient with stated confidence. The Wang cohort scanned at baseline, fraction 5, and fraction 15, so the clinical loop has two updates. Fraction 15 both corrects the fraction-5 estimate and re-prices the remaining budget.
The controller also prices the two glands separately. When the contralateral gland is the more sensitive one, dose can be shifted toward the already-compromised ipsilateral gland rather than spread evenly. That allocation decision has no counterpart in a single-threshold rule.
Research Program
The first stage is a go/no-go analysis on longitudinal cohort data. This means obtaining data from the Wang phase II cohort, or a matched cohort, under collaboration: ADC at baseline, fraction 5, and fraction 15, planning and delivered dose, and 6-month salivary outcomes. We fit a logistic model of 6-month xerostomia on delivered parotid dose alone. We then add ΔADC₅ after removing the part of it explained by delivered dose. If this residual adds no predictive value (likelihood-ratio test, cross-validated AUC gain), we stop. In that case we report that delivered-dose replanning is sufficient.
The second stage builds an in-silico adaptive testbed. The SegRap2023 nasopharyngeal carcinoma dataset (200 patients, 45 organ-at-risk contours plus tumor contours; SegRap2023) supplies anatomies. The open-source matRAD dose engine, as used by Text-to-Plan, generates intensity-modulated plans. Gland shrinkage and migration over the course are simulated to match the drift magnitudes reported in the delivered-dose study. Patient sensitivities are drawn from the stage-one posterior, and we check that the simulator reproduces the reported AUC of about 0.77 and the dose–ADC correlation of 0.60–0.68 before using it.
The third stage compares adaptation policies on identical simulated courses. The fourth is a retrospective replay on the real cohort. For each real patient we compute the plan each policy would have chosen and the predicted outcome. This replay is a necessary step before any prospective trial, but it does not substitute for one.
Baselines
The do-nothing reference is the original static plan delivered unchanged, scored on the simulated drifting anatomy. Its strengthened form adds the duct-sparing constraints of the Taizhou study, because that is the best published static approach to this exact toxicity. The second baseline is delivered-dose-triggered replanning. It accumulates dose on weekly cone-beam CT as in the delivered-dose study and replans to restore the original parotid objectives once accumulated dose exceeds the planned trajectory. This is the most direct competitor: it uses equipment every clinic already has, and the fraction-5 correlation suggests it captures much of the ADC signal.
The third baseline is the ADC threshold rule. Using the published cut-offs of 194.1 and 179.6 units, it replans with a fixed tightened parotid constraint for any gland that exceeds its threshold. The best reported evidence for the information it uses is AUC 0.772 (ipsilateral) and 0.766 (contralateral) for xerostomia prediction (Early prediction of parotid glands secretory function based on ADC variations during radiotherapy for nasopharyngeal carcinoma: a phase II prospective study). There is no reported result found on Lacuna for any adaptive intervention's effect on xerostomia rate, so no baseline outcome number exists to quote. For automated planning quality the nearest reported reference is FatPIN, which reached PTV D95 of 51.68 Gy against a 50.40 Gy clinical standard in 3–4 hours per plan, on cervical cancer rather than head and neck.
Two ablations isolate the proposal's ingredients. The first is a fixed-price adaptive planner that uses the same re-optimization with a constant $\lambda$ tuned on development patients. It tests whether updating the price matters. The second is our controller with sensitivity taken from delivered dose alone, with ADC removed. It tests whether ADC matters.
All policies share the same dose engine, the same anatomies, the same replanning opportunities (fractions 5 and 15), and the same tuning budget on a held-out development split. An oracle that knows each patient's true sensitivity sets the ceiling. Beating the field means two things. First, lower mean predicted xerostomia than both delivered-dose replanning and the threshold rule, at non-inferior PTV D95 (within 1%) and with no increase in brainstem, mandible, or oral-cavity limit violations. Second, a smaller fraction of patients whose individual budget is exceeded, with confidence intervals from patient-level bootstrap.
Why not just…?
Why not just replan from cone-beam CT delivered dose? This is the real threat. With r = 0.60–0.68 between ADC change and dose, delivered dose may carry most of the usable signal. The stage-one residual test settles this question directly, and the dose-only ablation settles it again in simulation. If the ADC residual adds nothing, we will recommend delivered-dose replanning and report the negative result.
Why not just apply the published ADC cut-off? A threshold makes a yes/no decision at one time point. It cannot tell how much to tighten, cannot trade one gland against the other, and cannot revise its decision at fraction 15. If the threshold rule matches our controller in stage three, the continuous posterior and the price updates are not worth their complexity, and we will say so. We expect any advantage to appear mainly in borderline patients near the cut-off and in asymmetric two-gland cases, and we will report results for those subgroups separately.
Why not just train a constrained reinforcement learner like QaSAL or an actor-critic planner? With two decision points per patient and no real closed-loop training data, a learned policy would have to be trained on our simulator and would inherit all of its errors. QaSAL also guarantees its constraints only on average over many episodes, which does not protect an individual patient. The fixed-price ablation shows how much is gained by updating the price. A learned policy could add something only beyond that, and we expect little.
Why not just use static duct sparing with wider margins? The duct-sparing authors themselves report that gland migration defeats static margins, and the delivered-dose study shows parotid V40 rising 44.4% above plan. Static duct sparing is our strengthened do-nothing baseline, so its shortfall is measured directly rather than assumed.
Decisive Checks
The project turns on three results. The first is whether ADC change, after removing the part explained by delivered dose, still predicts 6-month xerostomia in real data. The second is whether the controller beats delivered-dose replanning and the threshold rule in simulation at equal tumor coverage. The third is whether that gain survives the retrospective replay on real patients rather than appearing only in the simulator. A positive result on all three would justify a prospective feasibility study. A negative first result ends the method, and the report would argue that delivered-dose replanning is enough.
References
- Early prediction of parotid glands secretory function based on ADC variations during radiotherapy for nasopharyngeal carcinoma: a phase II prospective study
- Wireless Optimisation via Convex Bandits: Unlicensed LTE/WiFi Coexistence
- Parotid gland radiation dose‐xerostomia relationships based on actual delivered dose for nasopharyngeal carcinoma
- Parotid Gland Stem Cell Preservation during Intensity-Modulated Radiotherapy for Nasopharyngeal Carcinoma: Dosimetric Analysis and Feasibility
- Automatic Radiotherapy Treatment Planning with Deep Functional Reinforcement Learning
- Transforming Multimodal Models into Action Models for Radiotherapy
- QaSAL: QoS-aware State-Augmented Learnable Algorithms for Coexistence of 5G NR-U/Wi-Fi
- SegRap2023: A Benchmark of Organs-at-Risk and Gross Tumor Volume Segmentation for Radiotherapy Planning of Nasopharyngeal Carcinoma