Models that compare a patient's current scan with an earlier one are good at saying where something changed and poor at saying whether anything changed at all. When "no change" cases are added to a chest X-ray change-tracking task, accuracy falls to about 0.47 across every architecture tested (CheXRelNet). Foundation models on longitudinal brain MRI show the same pattern: they localize change with up to 97.3% accuracy but quantify it correctly less than 16% of the time, and often get the direction of change wrong (How Good are Foundation Models in Longitudinal MRI Disease Progression Reasoning?). We argue that "stable" should not be a learned category. It should be a statistical statement: the observed change is smaller than the change this model would report if the same patient were imaged twice in the same state.
Clinical measurement science already has a tool for this. Bland-Altman analysis plots the difference between two measurements against their mean, and summarizes agreement by the average difference (bias) and a 95% band around it (the limits of agreement). Applied to two measurements of an unchanged patient by the same method, the band describes that method's repeatability. We propose to estimate this band for a longitudinal model's own quantitative outputs, as a function of anatomical region, measurement size, and acquisition type (portable bedside versus standing films). The model then reports change only when its measured interval difference falls outside the band. This gives each progression call a number, a direction, and a stated noise floor. We then test whether this recovers the "no change" class that categorical models lose, especially when acquisition conditions shift.
Core Claim
A model that reports a continuous interval measurement, such as change in cardiothoracic ratio from anatomical boxes, will separate stable from changed exams more reliably under an empirically estimated, heteroscedastic repeatability band than under either a learned "stable" class or a single tuned threshold. "Heteroscedastic" here means the width of the band is allowed to depend on region, measurement size, and acquisition type. The claimed gain should be concentrated where noise differs across conditions: small structures, and pairs of films taken with different acquisition types. The same machinery gives a second, separate result. Bland-Altman agreement computed on interval deltas, not on single-timepoint values, is the correct way to report whether a model's progression measurements can stand in for a radiologist's. We expect it to disagree with level-wise agreement in informative ways, because errors that cancel between timepoints on one scan type add up across mixed types.
We deliberately do not propose a "Bland-Altman loss." Consider the mean squared error of paired differences $d_i = A_i - B_i$, where $A_i$ is the model's value, $B_i$ the reference value, $\bar d$ their average (the bias), and $s_d$ their standard deviation. It decomposes exactly as
$$\frac{1}{n}\sum_{i=1}^n d_i^2 = \bar d^{\,2} + \frac{n-1}{n}\, s_d^2 .$$
Minimizing mean squared error therefore already shrinks both the bias and the spread that set the limits of agreement. A differentiable $|\bar d| + 1.96\,s_d$ objective would be close to a reweighting of this quantity. The candidate gap listed in the research file (Gap 1) looks mostly illusory for that reason. The open problem lies elsewhere: deciding when a predicted change is real, which no training loss on level values answers.
How This Relates to Prior Work
Three longitudinal chest X-ray models motivate the gap and serve as baselines. BioViL-T splits image features into a static part and an interval-change part using a multi-image transformer. CoCa-CXR restricts cross-attention between the prior and current image to matching spatial neighborhoods. RECAP scores better/stable/worse on a disease-progression graph. All three treat stability as a class label and evaluate with accuracy or language metrics. None states how large a change must be before it counts.
CheXRelNet links the same anatomical region across two timepoints in a graph. It supplies the documented failure on "no change" that we target, and its region-level design is the natural backbone for our measurements.
The MRI foundation-model benchmark constrains the proposal. Its finding that models can localize but not quantify tells us to get quantities from explicit geometry (boxes or segmentations), not from free-text answers.
Bland-Altman practice supplies the method. The study of CT-predicted versus surgical peritoneal cancer index (Bland–Altman agreement analysis between CT predicted and surgical PCI) found Spearman correlation above 0.9 alongside limits from −5.46 to 6.32 points, with errors that depended on anatomical region. This supports modeling bands by region, not globally. Of prior work, the COVID-19 CT progression study (Longitudinal Quantitative Assessment of COVID-19 Infection Progression from Chest CTs) comes closest. It trains with a progression loss and checks volumetric agreement with Bland-Altman after the fact. It uses agreement only as a report card for two-timepoint lesion volumes, however, and never as the decision rule for calling change.
Research Program
The measurement layer takes anatomical boxes from a detector trained on Chest ImaGenome, which provides boxes for 29 anatomical locations and a manually verified gold subset from 500 patients. From these boxes it computes a small set of continuous quantities: cardiothoracic ratio, costophrenic-region lung-box height as a proxy for effusion, and a regional opacity score pooled from CheXRelNet-style region embeddings through a regression head. Radiologist reference values for the geometric quantities come from the gold boxes.
The repeatability layer is the core contribution. We collect "same-state" pairs: studies of the same patient in MIMIC-CXR acquired within a short window, filtered to reports that state no interval change. These pairs are split by patient into disjoint estimation and evaluation sets, so the band is never tuned on the pairs it is judged on.
On the estimation set we fit the conditional bias $b(m,g)$ and the empirical 2.5% and 97.5% quantiles $q_{lo}(m,g)$ and $q_{hi}(m,g)$ of the within-pair difference. Here $m$ is the mean of the two measurements and $g$ indexes region and acquisition-type combination. Following Bland-Altman practice, we condition on the pair mean and not on the first measurement. This avoids mathematical coupling, the artifact where one measurement's own noise appears on both axes and creates a spurious trend. Empirical quantiles replace the $\pm 1.96\,s_d$ rule because we expect heavy tails, particularly for portable films. The decision rule for a new pair with measured change $\hat\Delta$ is
$$\text{change} \iff \hat\Delta \notin \big[\, b(m,g) + q_{lo}(m,g),\; b(m,g) + q_{hi}(m,g) \,\big],$$
and the sign of $\hat\Delta$ gives the direction.
The analysis layer runs Bland-Altman on interval deltas against radiologist-derived deltas. It reports bias and limits separately for matched and mismatched acquisition types, so that level agreement and delta agreement can be compared directly.
As a secondary arm, we transfer the procedure to tumor volumes on Lumiere, drawn from the MRI benchmark cohorts. There we test whether the band raises Trend-F1 for segmentation-derived volume changes. This arm is conditional on having segmentations.
Baselines
The do-nothing reference predicts "stable" for every pair, which is the classification analogue of the last-context-image heuristic that CRONOS reports is hard to beat (NRMSE 7.55 for CRONOS against 8.38 for last-context-image on Lumiere). It sets the floor for stable-class recall and exposes how much of any "accuracy" is class prior.
The strongest existing categorical method on the stability question is CheXRelNet, which reports mean accuracy 0.68 and roughly 0.47 once "no change" cases are included, on Chest ImaGenome comparisons. On MS-CXR-T, CoCa-CXR reports 65.0% average accuracy across five conditions, 4.8 points above BioViL-T. We retrain CheXRelNet and CoCa-CXR on the same patient splits with the same backbone initialization and the same hyperparameter search budget as our regression head. We report stable-class recall, macro-F1, and sign accuracy.
Beating them means one of two outcomes. The first is higher stable-class recall at equal or better worsened/improved recall on MS-CXR-T and the Chest ImaGenome comparison relations. The second is matching macro-F1 overall while clearly beating them on mismatched-acquisition pairs.
Two ablations isolate the key ingredient. The first swaps the conditional band for a single global threshold on $|\hat\Delta|$, tuned on validation to maximize macro-F1. This is the fair version of "just tune the threshold," and it uses the same measurements. The second swaps empirical quantiles for the normal $\pm 1.96\,s_d$ band. If the global threshold matches the conditional band everywhere, the claim that noise depends on condition is false.
Why not just…?
Why not just rebalance the classes or tune the decision threshold of a categorical model? This is the most serious threat. A classifier's softmax margin is a continuous score, and a tuned cutoff might recover most of the stable class. The global-threshold ablation settles it. Our prediction is that a single threshold is too wide for large, repeatable quantities and too narrow for small, noisy ones. The conditional band should therefore win mainly on mismatched portable/standing pairs and small regions. If it does not, we will report that the gain comes from quantitative measurement alone, which is still a useful but smaller finding.
Why not just regress the change with mean squared error and read off correlation? By the decomposition above, mean squared error already controls bias and spread, and we use it to train the regression head. What it cannot supply is the stability decision: it gives a point estimate of change, not a statement of when that estimate is indistinguishable from repeat-imaging noise. Correlation hides systematic offsets entirely, as the peritoneal cancer index study shows (correlation above 0.9 despite limits wider than a surgically relevant margin).
Why not just use conformal prediction or dropout-based uncertainty on the predicted change? Our band is in fact a conditional split-conformal-style quantile, and we say so openly. The difference lies in what it is calibrated on: same-state pairs, which need only the label "no change," not ground-truth change magnitudes. Such magnitudes barely exist for chest films. We will add dropout-based predictive intervals as a further baseline. If they match the repeatability band on stable-class recall, the contribution reduces to the calibration data, and we will report it that way.
Decisive Checks
The central test is stable-class recall and macro-F1 on held-out MS-CXR-T pairs (1,326 image pairs) and Chest ImaGenome gold comparisons, stratified by acquisition-type match. We are looking for the conditional band to exceed both CheXRelNet's roughly 0.47 no-change accuracy and the tuned global threshold, most visibly on mismatched pairs.
The second check is a Bland-Altman plot of model-derived versus gold-box-derived changes in cardiothoracic ratio. It reports delta-wise bias and limits against level-wise ones. We predict that delta limits widen substantially on mismatched acquisitions even when level agreement looks acceptable.
The third check tests whether the band is honest. On the held-out same-state evaluation pairs, close to 95% of differences should fall inside it for every region and acquisition group. Systematic under-coverage in any group would falsify the calibration claim for that group.
The last check is Trend-F1 for Lumiere volume changes against the 0.19–0.63 range the MRI benchmark reports for foundation models. It shows whether the approach transfers beyond chest radiographs.
References
- CheXRelNet: An Anatomy-Aware Model for Tracking Longitudinal Relationships Between Chest X-Rays
- How Good are Foundation Models in Longitudinal MRI Disease Progression Reasoning?
- Learning to Exploit Temporal Structure for Biomedical Vision-Language Processing
- CoCa-CXR: Contrastive Captioners Learn Strong Temporal Structures for Chest X-Ray Vision-Language Understanding
- RECAP: Towards Precise Radiology Report Generation via Dynamic Disease Progression Reasoning
- Bland–Altman agreement analysis between CT predicted and surgical peritoneal cancer index in pseudomyxoma peritonei of appendiceal origin