Is It Really the Scanner? A Causal Audit of Cross-Device Performance Drops

Medical imaging benchmarks routinely report that a model trained on one manufacturer's scanner loses accuracy on another's, and they label this loss "hardware-induced domain shift." In almost every public multi-device benchmark, though, the device is not the only thing that changes. Different hospitals own different scanners, so the switch to a new device also brings a different patient population, different operators, different annotation habits, and often a different preprocessing pipeline. The headline drop is therefore an observational number: it is correlated with the change of device, but nobody has shown how much of it the device causes. This proposal builds an audit protocol that splits a reported cross-device drop into the part caused by the device and the parts caused by population, labeling, and evaluation configuration. It then asks whether the domain-adaptation methods that "win" these benchmarks still win once only the device-caused part is counted.

The plan rests on one rare resource: paired acquisitions, where the same eye is imaged on two devices. In a paired dataset the anatomy is held fixed and only the hardware changes, so this is as close to a real intervention on the device as clinical data allows. We use paired data as ground truth to validate cheaper simulated device interventions. We then carry those validated interventions to unpaired multi-vendor benchmarks, where device and site cannot otherwise be separated. Together these steps turn "cross-device generalization" into a quantity that can be reproduced and causally interpreted.

Core Claim

Our claim is that cross-device benchmark gaps, as currently reported, mix up the causal effect of acquisition hardware with site, population, and pipeline effects. We expect a substantial and benchmark-dependent share of the reported gap not to be hardware at all. Rankings of adaptation methods computed on the total gap can differ from rankings computed on the device-attributable gap.

Concretely, let $m$ be a task metric, $S$ the source domain and $T$ the target domain. We write the reported gap as

$$\Delta_{\text{total}} = \Delta_{\text{device}} + \Delta_{\text{population}} + \Delta_{\text{pipeline}} + \Delta_{\text{interaction}},$$

where $\Delta_{\text{device}}$ is the change in $m$ when only the acquisition device is switched for the same subjects, $\Delta_{\text{population}}$ is the change from the different case mix and labels at the target site, $\Delta_{\text{pipeline}}$ is the change from preprocessing and inference settings, and $\Delta_{\text{interaction}}$ is the remainder. Benchmarks today report only the left-hand side. We estimate each term on the right.

Where This Sits in Prior Work

Three lines of work motivate the proposal by exposing the gap.

Causality is Key for Interpretability Claims to Generalise audits interpretability papers against Pearl's ladder, the three levels of causal evidence: association, intervention, and counterfactual. It finds that 53.5% of 186 claims overstate their causal level. We apply the same lens to a different object. A cross-device benchmark gap computed from unpaired sites is first-rung evidence, yet it is routinely described in second-rung language ("the device causes the drop").

Beyond Benchmarks: Toward Causally Faithful Evaluation of Large Language Models supplies the method for the pipeline term. It treats evaluation settings (prompt format, shot count, decoding parameters) as factors in a factorial experiment, which means running every combination of settings and using analysis of variance to measure each factor's effect. It reports accuracy swings of up to 70% and ranking reversals in 9 of 18 pairwise comparisons on AIME '24. That work only touches text inputs and software knobs. We carry its factorial design over to image preprocessing and detector inference settings, and we add the population and device terms it never faces.

Benchmarks as Software: A Case Study on Terminal-Bench shows that much of what looks like a capability gap can be harness artifact. A 29.9-point spread between harnesses largely disappeared when the timeout was scaled 16x. That is the reproducibility analogue of our pipeline term.

Two papers constrain the design. The first is Optical Coherence Tomography Harmonization with Anatomy-Guided Latent Metric Schrödinger Bridges, which translates OCT scans between scanner brands. It shows that unconstrained generative translation moves anatomical boundaries: retinal boundary error was 0.89 pixels for DDIB versus 0.43 for their anatomy-guided method. The second is AI Alignment in Medical Imaging: Unveiling Hidden Biases Through Counterfactual Analysis, which uses diffusion-generated counterfactual X-rays to test whether predictions stay the same when demographic attributes change. Both warn that a synthetic "change the device" image may also change the anatomy. For that reason we never trust a simulated device intervention until it has been checked against real paired data.

Two papers supply the data for the device term:

The adversarial and causal-masking methods are our objects of study, not our method:

Both claim robustness to acquisition differences. Neither shows that the improvement comes from the device term rather than the population term.

Research Program

The program has three stages, and each produces a reusable artifact.

Stage 1: paired ground truth. On AI-READI we compute the device effect directly. Each participant has a Triton and a Maestro2 image, so a model's prediction difference between the two images of the same eye is a per-subject device effect. Averaging these differences gives $\Delta_{\text{device}}$ with no population confounding. We do this for several fixed pipelines:

  • frozen RETFound embeddings feeding biomarker regressors;
  • AutoMorph vessel biomarkers;
  • a fine-tuned classifier on whichever participant-level label AI-READI provides.

Stage 2: validating device simulators. We build a ladder of increasingly expressive interventions that turn a source-device image into a plausible target-device image:

  • per-channel color and illumination matching, plus blur and noise matched to target-device statistics;
  • a learned translation network trained on unpaired images;
  • an anatomy-guided translation in the spirit of the latent Schrödinger bridge.

Each simulator is scored by whether it reproduces the real paired effect at the level of individual images. The test is the correlation between the model's prediction change under the simulator and its prediction change between the two real images, together with the gap between simulated and real $\Delta_{\text{device}}$. A simulator passes only if it reproduces the device effect and also leaves vessel geometry intact. The geometry check uses AutoMorph tortuosity and caliber, because the feature-translation paper reports that fine geometric biomarkers transfer worst. This stage directly answers whether a "hardware mirror" can be trusted. The software analogue, StableToolBench-MirrorAPI, validated its simulated APIs against real ones, and we hold device simulators to the same kind of check.

Stage 3: decomposing unpaired benchmarks. On FUSEP, on the multi-scanner OCT data, and on the multi-vendor 3D echocardiography data, there are no paired images. Here we estimate $\Delta_{\text{device}}$ by applying the validated simulator class to source-site images. These are source patients, imaged virtually on the target device, scored against source labels. Comparing three quantities separates the terms:

  • the simulated-device score against the in-domain score, which isolates the device term;
  • the simulated-device score against the true target-site score, which leaves population plus label effects;
  • a factorial sweep over preprocessing settings (resize policy, intensity normalization, input resolution, confidence threshold, non-maximum suppression threshold), which gives the pipeline term in the style of Beyond Benchmarks.

We treat simulator validity as transferable only within a modality family. Fundus results justify the color, illumination, and blur class. For ultrasound and OCT we fall back on whatever internal anchors exist, and if none exist we report the device term as bounded rather than identified.

This decomposition can fail in an informative way. If in FUSEP each vendor appears at only one center, then device and site are fully aliased in the real data, and the only handle on the device term is the simulator. We will say so explicitly. That would itself be a finding about what multi-vendor benchmarks can and cannot claim.

Baselines

The naive reference point is the reported total gap: in-domain score minus target-site score under the benchmark's default pipeline, with all of the gap attributed to hardware. This is what FUSEP and the cardiac and OCT papers implicitly report. For FUSEP, the file gives Relation-DETR as the top detector at 85.6% mean average precision (mAP) on crown–rump-length views and 95.6% on nuchal-translucency views. The file does not break these numbers down by cross-vendor split, so we will recompute the per-vendor gaps under the released protocol before decomposing them. We beat this reference if the decomposition attributes a statistically significant share of the gap, with confidence intervals from bootstrapping subjects, to non-device terms on at least one benchmark.

The strongest competitor for the attribution question is covariate adjustment on recorded metadata. This means stratifying or reweighting target-site cases to match source-site gestational age, view type, image quality and similar variables, then recomputing the gap and calling the residual "device." It is cheap and standard, and if the metadata is rich it could recover most of what we recover. It runs on exactly the same splits, models, and metric. On AI-READI we can score it against paired truth: we deliberately build unpaired splits in which device is confounded with a recorded participant attribute, then check whether metadata adjustment recovers the paired $\Delta_{\text{device}}$. We beat it if our simulator-based estimate is closer to paired truth under these induced-confounding splits.

The strongest existing adaptation methods are our objects of measurement:

  • C3, which reports an average 8.5% Dice improvement over V-Net and a mean surface distance reduced from 3.56 mm to 2.01 mm on Philips EPIQ 7C to GE Vivid E9;
  • ToMo-UDA as used in FUSEP, for which no reported result was found on Lacuna;
  • CausalCLIPSeg, which reports 85.21% Dice on QaTa-COV19 versus 83.66% for LViT, with a 1.5% Dice drop when its causal module is removed.

Each method is retrained with the same backbone, the same augmentation, and an equal hyperparameter search budget (a fixed number of trials per method). We then re-rank the methods by how much of $\Delta_{\text{device}}$ versus $\Delta_{\text{total}}$ they recover.

The ablation that isolates our key ingredient replaces the validated simulator with an unvalidated one: the unconstrained translation network accepted without the Stage 2 check. If the decomposition is equally accurate against paired truth without validation, then paired validation is not doing the work, and the proposal reduces to "use a translation model."

Why not just…?

Why not just report per-site results and adjust for metadata? This is the real threat. It costs almost nothing, and on benchmarks with rich acquisition metadata it may be close to sufficient. Its weakness is that it can only adjust for what was recorded. Operator technique, referral patterns, and annotator conventions usually are not recorded, and these are exactly the factors that move with the site. The induced-confounding test on AI-READI settles this empirically. If metadata adjustment matches paired truth as well as our method does, we will recommend it and drop the simulator stage for well-annotated benchmarks.

Why not just pool multi-vendor training data? Pooling may reduce the total gap, but it does not tell anyone which part of the gap was hardware, which is the question here. Evidence in the file also suggests pooling does not remove acquisition shortcuts. The AI Alignment paper reports models trained on large multi-center chest X-ray cohorts that pass standard accuracy checks while failing counterfactual invariance tests. We include pooled training as one of the re-ranked methods, so its share of the device term is measured rather than assumed.

Why not just harmonize the images and see what drop remains? Harmonization changes the inputs. It does not measure the effect of the device on the inputs, and a harmonizer that also alters anatomy will under- or over-state the device effect in ways nobody can detect without paired data. The OCT bridge paper's 0.89-versus-0.43-pixel boundary error is direct evidence of that risk. In our program, harmonizers enter as candidate simulators and must pass the paired check.

Why not just run the factorial pipeline sweep from Beyond Benchmarks? That captures $\Delta_{\text{pipeline}}$ only. It treats the data as fixed, so it cannot separate device from population. We run the sweep as one component. If the pipeline term alone explains most of the gap, as the Terminal-Bench timeout result might lead one to expect, that is a reportable outcome and makes the rest of the program cheaper.

Decisive Checks

The first check is on AI-READI. The simulator-based estimate of $\Delta_{\text{device}}$ must fall inside the bootstrap confidence interval of the paired estimate, and its per-image prediction changes must correlate with the real paired changes. We fix the correlation threshold before running and report it. If no simulator class passes, the program stops after Stage 2. Its contribution then becomes the negative result that synthetic device interventions are not yet reliable enough to attribute benchmark gaps.

The second check is the induced-confounding test. Our estimator must stay near paired truth as the confounding between device and participant attribute is increased, while the naive gap drifts away.

The third check is the substantive one. On FUSEP and the cardiac ultrasound benchmark, we compute the Spearman rank correlation between adaptation-method rankings on $\Delta_{\text{total}}$ and on $\Delta_{\text{device}}$. For comparison, Terminal-Bench reports $\rho = 0.958$ for model rankings before and after task fixes.

The claim fails if rankings are essentially preserved and the device term accounts for nearly all of the gap. In that case, the field's current practice is vindicated, and we have a reusable protocol showing why. A low correlation, or a large non-device share, would mean that published cross-device robustness claims need restating as site-robustness claims.