Dosimetric Disagreement: Measuring Which Contouring Differences Matter With a Fast Dose Surrogate

When two radiation oncologists outline the same tumor or organ on a CT scan, their outlines differ, sometimes by several millimeters. Research on this disagreement treats it as a geometry problem: models learn to reproduce the spread of expert outlines and are scored on overlap measures such as Dice. In the clinic, though, an outline matters only through the radiation dose it leads to. A 2 mm disagreement at the edge of the esophagus next to a high-dose target can change organ dose a great deal. The same disagreement in the middle of the bladder may change almost nothing. We propose to measure contouring disagreement in units of dose. A neural dose surrogate, a network that predicts in one fast pass the 3D dose a treatment plan would produce from a CT and its outlines, is used to ask a specific question: if the plan is built on rater A's outline, how much does the dose that rater B's version of the structure receives deviate from what the plan intended?

This question is not the same as feeding each sampled outline through a dose predictor and looking at the spread of outputs. A dose predictor learns that the planned dose wraps around whatever target it is given, so every sample looks well covered on its own terms, and the actual risk stays hidden. The risk appears only when we plan on one outline and evaluate on another. We build that cross-evaluation, check that a fast surrogate reproduces it faithfully against a real planning system, and test whether it ranks disagreements differently, and more usefully, than geometric metrics or cheap distance-based proxies.

Core Claim

Our claim is that a fast, differentiable dose surrogate turns inter-observer variability into a dosimetric quantity that geometric metrics do not capture. Let $s_1,\dots,s_K$ be $K$ outlines of a patient's structures, from real raters or sampled from a multi-rater model. Let $D(s)$ be the surrogate's predicted dose when planning on outline set $s$, and let $m(D, s)$ be a clinical dose-volume metric of dose $D$ evaluated on structure $s$, for example the dose covering 95% of the target or the mean dose to an organ at risk. We define the cross-evaluated dosimetric disagreement as

$$\Delta_{ij} = m\big(D(s_i), s_j\big) - m\big(D(s_i), s_i\big),$$

which is the error in metric $m$ when the plan built on outline $i$ is judged against outline $j$. The full $K\times K$ matrix of $\Delta_{ij}$ per structure is the object of study.

The claim is falsifiable in two ways. First, the surrogate must track how a real planning system responds to outline changes. Second, the matrix must rank cases and structures differently from Dice, Hausdorff distance, or distance-to-target-weighted geometric error. Only then does it justify using a neural surrogate at all.

Relation to Prior Work

Two literatures meet here, and neither has crossed into the other. Neural dose surrogates take outlines as fixed inputs. DiffDP predicts 3D dose with a diffusion model conditioned on CT and outlines, avoiding the blurring of regression models but needing 1,000 denoising steps. Flexible-Cm GAN adds a differentiable dose-volume loss and supports interactive beam changes. LENAS distills an ensemble of searched architectures into a single fast student on OpenKBP. Chen et al. add a distance-to-target map to a ResNet predictor. All four supply the method we reuse: a single-pass, outline-conditioned dose predictor. None of them asks how its output changes when the outline changes.

Physics-Guided Radiotherapy Treatment Planning shows that a pre-trained differentiable dose predictor can carry useful gradients back into an upstream network. That result supports the second half of our program, but that work still treats masks as fixed truth.

On the other side, multi-rater models reproduce expert spread: Pionono with per-rater latent distributions, ProSeg by separating image ambiguity from rater preference, MoSE with a mixture of stochastic experts trained by optimal transport, and MRNet with rater-agreement modules. Benchmarks like QUBIQ score this spread only geometrically.

These works motivate the gap: they give us a way to generate plausible outlines but no way to say which differences matter. The OpenKBP finding that voxel-wise dose accuracy and dose-volume scores do not move together is a caution as well as a motivation. It means our evaluation must be stated in dose-volume metrics, not voxel error.

Research Program

The first component is a fidelity study, and the project depends on it. Dose predictors are trained on one outline per patient and may largely ignore small outline edits, leaning instead on CT appearance. That would make them useless for this purpose; the research file lists this sensitivity as uncharacterized. We train a single-pass predictor on OpenKBP (a U-Net and a LENAS-style distilled student). We then apply controlled outline perturbations: local dilations and erosions, and rater-style displacements. For a subset of cases we replan each perturbed outline set in a clinical treatment planning system with a fixed automated protocol. This is the expensive step and requires a research planning system license. We measure whether the surrogate's predicted change in each dose-volume metric correlates with the replanned change. If fidelity is poor, we retrain with perturbation augmentation, using these replans as extra training targets, and report both versions.

The second component is real multi-rater data in a dose context. The file contains no dataset with both multiple raters and dose, so we will build one. Four to five clinicians will re-contour the primary target and the most sensitive organs at risk (brainstem, spinal cord, parotids, optic structures where present) on roughly 30 OpenKBP test cases. This gives real $\Delta_{ij}$ matrices from both the planning system and the surrogate.

The third component is the analysis that tests the claim. We rank (case, structure) pairs by surrogate $\Delta$, by geometric disagreement, and by a distance-weighted geometric proxy. We then measure how well each ranking predicts the planning-system $\Delta$.

The fourth component is secondary and conditional on the first three succeeding. We fine-tune a multi-rater model (Pionono or ProSeg architecture) on the re-contoured cases with an added loss. The loss is the one-dimensional Wasserstein distance between the distribution of $\Delta_{ij}$ over model samples and over real raters, back-propagated through the frozen surrogate using soft masks. The aim is a model whose spread is correct where it matters for dose, not just on average. We also test the surrogate's use as a review tool, flagging the regions where rater disagreement drives the largest $\Delta$.

Baselines

The naive reference is no propagation: report Dice, Hausdorff distance, and generalized energy distance between raters, as current multi-rater work does, and treat high geometric disagreement as high risk. Its ranking is computed on the same re-contoured cases and compared against planning-system $\Delta$ with the same rank-correlation and top-decile-recall metrics.

The most dangerous competitor is a distance-weighted geometric proxy. It weights each voxel of disagreement by its proximity to the high-dose target, using the distance-to-target map of Chen et al., and then sums. This proxy needs no dose model and could capture most of the effect. We tune its single free parameter, the distance decay scale, on held-out cases with the same budget we use for any surrogate choice. The surrogate must beat it by a margin outside the bootstrap interval on rank correlation with planning-system $\Delta$ to justify itself.

The strongest existing method for the surrogate itself is the LENAS ensemble. LENAS reports an OpenKBP Dose Score (mean absolute error) of 2.357 for the ensemble, 2.565 for the distilled student, and 2.619 for a baseline U-Net. Our predictor must match the student-level score, within about 0.1, on the unperturbed test set; otherwise gains under perturbation could reflect a weaker model rather than a better design.

For the fourth component, the baselines are the unmodified Pionono and ProSeg models trained on the same cases with the same epochs and tuning budget. No result on our new data exists for them; ProSeg reports a generalized energy distance of 0.1152 on LIDC-IDRI and a 20.71% improvement in generalized energy distance on its NPC multi-rater set, but neither involves dose. Success means lower Wasserstein error on $\Delta$ distributions with no more than a small loss in generalized energy distance.

The key ablation replaces cross-evaluation ($i\neq j$) with self-evaluation (each sample's own dose on its own outline). This isolates whether the plan-on-one, judge-on-another construction is what carries the signal.

Why not just…?

Why not just run the clinical planning system on every outline? For evaluation this is the gold standard, and we use it as ground truth on a subset. Automated replanning takes minutes per outline set, as the physics-guided planner paper notes for classical inverse planning. The full program needs a $K\times K$ matrix per structure per patient, sampled outlines for training, and gradients for the fourth component. A planning system provides none of these at scale and no gradients. If the surrogate fails the fidelity study even after augmentation, this alternative wins and we will say so.

Why not just weight geometric disagreement by distance to the target? This is a real threat. Dose falls off mostly with distance from the target, so the proxy may explain most of the variation in $\Delta$. It cannot capture beam-path effects, organs that compete for sparing, or the asymmetry between planning on a larger versus smaller target, which $\Delta_{ij}$ encodes and a symmetric distance weight does not. The head-to-head ranking experiment settles this directly. If the proxy matches the surrogate, the honest conclusion is that a cheap proxy suffices, and the proposal reduces to recommending it.

Why not just add a fixed safety margin? Isotropic margins are the clinical answer to uncertainty, but the evidence compiled from SegTHOR and You et al. indicates they break down where organs at risk border the target. We include a margin condition: plan on the consensus outline expanded by 3 mm, then evaluate against each rater. This shows whether margins already absorb the measured $\Delta$.

Why not just push sampled outlines through the dose predictor and report output spread? This is the obvious combination, and it is our self-evaluation ablation. Because the predictor learns that dose conforms to whatever target it receives, we expect it to understate risk. The ablation quantifies by how much.

Decisive Checks

Three results decide the project. First, on the perturbation replans, the surrogate's predicted change in target 95% coverage and organ mean dose must correlate strongly with planning-system changes. A rank correlation below roughly 0.7 after augmentation would mean the surrogate is not fit for propagating contour uncertainty. That would itself be a useful negative finding for the surrogate literature. Second, on the re-contoured OpenKBP cases, the surrogate $\Delta$ ranking must beat both geometric metrics and the tuned distance proxy at predicting planning-system $\Delta$, with bootstrap intervals over patients. Cross-evaluation must also beat self-evaluation. Third, only if the first two hold, dose-aware fine-tuning must bring the model's $\Delta$ distribution closer to the raters' distribution than plain Pionono or ProSeg, under five-fold cross-validation on the re-contoured set, with inference cost reported per sample, since the surrogate is single-pass and DiffDP-style sampling is ruled out by latency.