When several experts outline the same brain structure, their outlines differ. Recent models learn this spread directly, giving each annotator a small latent "style" vector and sampling plausible alternative segmentations. These models assume all of the spread comes from the people. In MRI, though, part of what makes a boundary hard to see is the scanner. Receiver coils are sensitive in a smooth, spatially uneven way. This produces a bias field, a slow multiplicative shading over the image, and in deep or far-from-coil regions the image is both darker and noisier. If annotators disagree more in those regions, a model of annotator style can quietly absorb the coil's sensitivity map. It will then report "experts disagree here" when the honest statement is "this scanner measured this region poorly."
We propose to test this confound and fix it in the setting where it matters most and is easiest to isolate: segmenting white matter tracts from high-angular-resolution diffusion imaging (HARDI). HARDI is an MRI protocol that measures water diffusion along dozens to hundreds of directions, so the model can resolve fiber bundles that cross inside a voxel. Diffusion data have a useful physical property here. Dividing each diffusion-weighted image by the non-diffusion-weighted reference cancels the coil shading exactly, but it leaves the noise untouched. Shading therefore never shows up as a brightness difference in the features a tract segmenter sees. It shows up only as spatially varying reliability. The core proposal is a probabilistic tract segmenter with two separate uncertainty terms: one attributable to annotators, and one tied explicitly to a physically estimated local signal-to-noise map. We train it with a supervision trick: degraded re-acquisitions that by construction share identical labels. We then test whether this separation makes uncertainty transfer from research scanners to clinical acquisitions.
Core Claim
Multi-rater uncertainty models will misattribute acquisition-driven ambiguity to human disagreement unless acquisition quality is given its own channel. In diffusion MRI that channel can be computed from physics rather than learned. The observed diffusion-weighted signal along gradient direction $k$ at location $x$ is $S_k(x) = B(x)X_k(x) + N(x)$. Here $B(x)$ is the smooth coil bias field, $X_k(x)$ is the true signal, and $N(x)$ is noise of fixed scale $\sigma$. The normalized signal $S_k/S_0$ removes $B$, but its noise does not cancel. The quantity that governs how reliable each voxel's fiber estimate is becomes
$$s(x) = \frac{B(x)\,X_0(x)}{\sigma},$$
the effective signal-to-noise ratio, where $X_0$ is the true non-diffusion-weighted signal. Our model's per-voxel logit for annotator $r$ is the sum of a shared anatomical prediction, an annotator term driven by a latent $z_r$, and a zero-mean perturbation whose variance is a learned but monotonically decreasing function of $s(x)$ and of the number of gradient directions. The annotator term is blocked from seeing $s(x)$, and the acquisition term is blocked from seeing annotator identity.
The new claim is not "model rater variability," which Pionono and Liu et al. already do. Nor is it "correct the bias field," which S2DNets and NCBC do. The claim is that each kind of uncertainty can be supervised so that it means what it says, and that this attribution is what lets uncertainty generalize across acquisition protocols.
Positioning Against Prior Work
The rater-modeling papers supply the method we build on. Pionono gives each annotator a Gaussian latent whose mean encodes systematic bias and whose variance encodes self-inconsistency. It is validated on 2D histopathology, where coil shading does not exist. Liu et al. sample a six-dimensional latent to produce multiple hypotheses on single-contrast brain MRI and treat image intensities as clean. The QUBIQ challenge found that plain ensembles beat single variational networks at matching multi-rater distributions. That result constrains us: any single-network design must be compared against an ensemble, not against a weak variational baseline. None of these papers asks whether the learned annotator variance tracks scanner sensitivity, and none touches diffusion data.
The bias-field papers motivate the "do not just divide it out" stance. NCBC shows on endorectal diffusion-weighted MRI that correcting bias and removing noise as separate steps distorts the noise statistics. Dividing by a small bias factor inflates noise where the coil is weak. We take this as evidence that bias correction cannot recover lost information, only redistribute it. That is why we represent reliability rather than correct intensity. S2DNets and the adversarial bias augmentation of Chen et al. (described in the file) push segmenters toward invariance to shading. Invariance is the right goal for the mean prediction and the wrong goal for the uncertainty.
On the diffusion side, TractSeg supplies the backbone. It feeds three fiber-orientation peaks per voxel to convolutional networks and segments 72 tracts deterministically on preprocessed Human Connectome Project data. It struggles with thin tracts such as the fornix, which sits deep in the brain where coil sensitivity is lowest; we read this as a hint of the confound we target. The QBI repeatability study is the only file entry measuring human operator variability for tracts. It finds that higher angular resolution reduces operator dependence near lesions, which supports our premise that acquisition quality and human variability are entangled. The method it studies is manual seeding rather than learned segmentation.
For cross-protocol generalization, the gradient-map approach removes scalar intensity style with edge maps. That discards orientation and does not apply to fiber peaks. Anatomy-to-tract mapping bypasses diffusion entirely and cannot see microstructural change. We treat "modality" narrowly and honestly: the shift is between a research protocol (HCP, many directions, three shells, a high-sensitivity head coil) and clinical-style multi-site protocols (TractoInferno). This is a shift in angular sampling and spatial noise profile within one physical measurement, not a jump such as CT to MRI.
Research Program
The work has three parts.
Measure the confound where labels already exist. On the QUBIQ brain-growth task, which has seven raters, we estimate each scan's bias field. We then check whether real inter-rater disagreement concentrates in low-sensitivity regions after controlling for distance to the boundary and local tissue contrast. Next we inject synthetic smooth bias fields with matched noise into held-out scans and retrain or fine-tune the Liu et al. and Pionono-style models. If their per-annotator spread rises in the shaded regions even though the labels are unchanged, the models are absorbing coil effects into "human" variance.
Build multi-rater tract data, which do not exist in the file. We will collect delineations from four to six trained annotators for the corticospinal tract, optic radiation, arcuate fasciculus, and fornix. The work covers about 40 HCP subjects and 30 TractoInferno subjects, starting from TractSeg-style candidate bundles that annotators curate. A subset is re-annotated after a delay to measure each annotator's self-consistency. This is the most expensive part of the project and the main risk.
Train the decomposed model with a label-sharing supervision signal. From each HCP subject's raw data we synthesize several degraded re-acquisitions. Each one applies a realistic coil sensitivity profile drawn from a library of fitted fields, adds Rician noise, and subsamples gradient directions down to clinical-like single-shell schemes of 30 to 64 directions. We recompute the fiber peaks and the map $s(x)$ for each copy. All copies of a subject share the same annotator labels. Any prediction spread across copies is therefore acquisition-driven by construction and is assigned to the acquisition term. Spread across annotators on a fixed acquisition is assigned to the annotator term.
We estimate $s(x)$ from the reference image using a bias estimate (N4 or DeepN4) together with a background or repeated-$b{=}0$ noise estimate. Because $s(x)$ is computed from physics, it remains available at test time on any scanner without labels or test-time optimization. That avoids the latency and class-prior fragility of PITTA.
Baselines
The naive reference is TractSeg trained on the majority-vote or STAPLE consensus of our annotators, with N4 applied to the reference image. It produces no uncertainty except softmax confidence. It anchors segmentation accuracy and shows how poorly raw confidence tracks either kind of disagreement. TractSeg reports gains of 9 Dice points over TRACULA and RecoBundles on HCP. The file gives no absolute tract Dice, and there is no reported result found on Lacuna for multi-rater tract segmentation uncertainty. Our baselines are therefore re-run, not quoted.
The strongest competitor for matching the annotator distribution is a five-member deep ensemble of TractSeg models. Each member is trained on a different annotator subset, as in the winning QUBIQ entries, which reached average Q-Dice between 0.80 and 0.85 QUBIQ. We give it the same bias-and-noise augmentation as our model, which makes it the robust-ensemble alternative a reviewer would propose.
The strongest single-network competitor is a Pionono-style annotator-latent model on the TractSeg backbone. For reference, Pionono reports Cohen's κ of 0.758 on Gleason 2019 Pionono, and Liu et al. report continuous Dice of 0.5372 on QUBIQ brain growth Liu et al.. We reproduce both numbers on their original data before porting the methods, so that implementation errors do not flatter us.
The key ablation feeds $s(x)$ to the annotator-latent model as an extra input channel without the separated acquisition term and without label-sharing supervision. If this matches the full model, the decomposition is unnecessary.
All methods use the same subjects, splits, augmentation library, backbone, and number of tuning trials. The ensemble receives five times our training compute, and we report its inference cost.
We count it as beating the baselines if three things hold. First, we match the ensemble's Q-Dice and continuous Dice against held-out annotators within noise. Second, the annotator-variance term stays invariant under injected degradation, meaning it does not correlate with $s(x)$, while both the ensemble and the plain latent model show significant leakage. Third, the acquisition term ranks segmentation errors on TractoInferno better than ensemble variance does.
Why not just…?
Why not just run N4 or DeepN4 first? In normalized diffusion data the bias field is already divided out, so intensity correction changes almost nothing that the segmenter sees. What it cannot change is the reduced SNR in weakly sensed regions. NCBC shows that dividing out the bias inflates noise there. N4 is therefore present in every arm of our experiments, and the injected-degradation test shows directly whether N4-preprocessed models still leak acquisition variance into annotator variance.
Why not just ensemble with heavy bias and noise augmentation? This is the real threat. Augmented ensembles may well produce variance that rises in low-SNR regions and may match the annotator distribution well, as QUBIQ suggests. But ensemble variance is a single number per voxel. It cannot say whether the spread came from people or from the scanner, and that distinction decides whether a surgeon should re-scan or consult a colleague. Our evaluation tests this attribution explicitly. If the augmented ensemble's errors on TractoInferno are predicted as well by its variance as by our acquisition term, and if a post-hoc regression of its variance on $s(x)$ recovers an equally clean split, we will report that the decomposition adds interpretability but not accuracy.
Why not just add the SNR map as an input channel? This is our main ablation. Without labels shared across degraded copies, nothing forces the network to route SNR-driven spread away from the annotator latent. We expect leakage to persist. If it does not, the simpler model wins and we adopt it.
Why not just reach consensus with STAPLE? STAPLE removes the annotator signal we want to keep. Pionono and Liu et al. already show that consensus training loses alternative hypotheses, so it serves only as the naive anchor.
Decisive Checks
We consider four tests decisive.
The first is the QUBIQ leakage test. If injected shading does not shift annotator variance in existing latent models, the motivating confound is weak in structural MRI, and we narrow the claim to diffusion data.
The second is a stratified analysis on our multi-rater tract labels. We measure rater disagreement against $s(x)$ separately for the fornix and the corticospinal tract. The fornix is predicted to show the strongest acquisition share.
The third is a transfer test. We train only on HCP plus synthetic degradations, then evaluate on TractoInferno sites unseen during training. We report Dice and Q-Dice against held-out annotators, the area under the curve for detecting voxel errors from each uncertainty term, calibration error, and correlation between annotator variance and $s(x)$. The proposal succeeds if the acquisition term predicts cross-site errors better than ensemble variance does while annotator variance stays stable across sites.
The fourth is a sanity check on real repeat acquisitions, if they can be obtained. Disagreement between scans of the same subject should be explained by the acquisition term, not by the annotator term.
References
- Probabilistic Modeling of Inter- and Intra-observer Variability in Medical Image Segmentation
- Variational inference for quantifying inter-observer variability in segmentation of anatomical structures
- Structure and Smoothness Constrained Dual Networks for MR Bias Field Correction
- Noise-Compensated, Bias-Corrected Diffusion Weighted Endorectal Magnetic Resonance Imaging via a Stochastically Fully-Connected Joint Conditional Random Field Model
- QUBIQ: Uncertainty Quantification for Biomedical Image Segmentation Challenge
- TractSeg - Fast and accurate white matter tract segmentation
- Anatomy-to-tract mapping infers white matter pathways without diffusion streamline propagation